Instruction Manual
11-1
IM 01R04B04-00E-E 8th edition March 01, 2011 -00
All Rights Reserved. Copyright © 2003, Rota Yokogawa
Qnom is the water ow rate at about 1 bar pressure drop.
The owmeter has a default low cut of 0.05% of Qnom.
Accuracy mass ow
Liquid : ± 0.1% of ow rate
± zero stability / ow rate *100%
(refer to table 11-2)
Gas (option /GA) : ± 0.5% of ow rate
± zero stability / ow rate *100%
(refer to table 11-2)
Accuracy volume ow
SQRT ( (mass ow error in %)² + (density error in %)²)
Please refer to the Yokogawa sizing software.
Accuracy based on the frequency output includes the com-
bined
effects of repeatability, linearity and hysteresis.
Repeatability for liquids: ± 0.05%
± (zero stability/2) / ow rate*100%
Table 11-2 : Zero Stability
Type RCCS30 RCCS31 RCCS32 RCCS33
kg/h 0.005 0.0085 0.019 0.045
Type RCCx34 RCCx36 RCCx38 RCCx39
RCCx39
/IR
RCCx39
/XR
kg/h 0.15 0.5 1.6 5 13 25
- Remote detector RCCS30 to 33: 2 tubes, low ow design
- Remote detector RCCS34 to 39/XR: 2 tube design
- Remote eld-mount converter RCCF31
- Remote rack-mount converter RCCR31
- Integral type RCCT34 to 39/XR: 2 tube integral design
: Liquid, gas or slurries
: Mass ow, density, temperature
and derived from these values:
concentration, volume ow and
net ow
Table 11-1: measuring range
Type RCCS30 RCCS31 RCCS32 RCCS33
Qmax t/h 0.1 0.3 0.6 1.5
Qnom t/h 0.045 0.17 0.37 0.9
Type RCCx34 RCCx36 RCCx38 RCCx39
RCCx39
/IR
RCCx39
/XR
Qmax t/h 5 17 50 170 300 600
Qnom t/h 3 10 32 100 250 500
The stiffness of the ROTAMASS tubes is slightly line pressure
dependent. The static pressure effect of mass ow and
density
can be corrected by setting the static pressure manually
via menu.
Table 11-3 : Static pressure effect on mass ow (if not corrected)
---- ---- ---- ---- 0.00081
0.00000 0.00012 0.00246 0.02094 0.00084
0.00346 0.00950 0.01058 0.02920 0.00740
0.00336 0.00896 0.00808 0.01780 ----
Adjustment with water and air at calibration temperature.
With option /K4 thermal stabilized.
For option /K6 see also “Special calibrations” on page 11-2.
Measuring range: 0.3 kg/l to 5 kg/l (RCCx39,
RCCx39/IR and RCCx39/XR to 2 kg/l)
No density measurement for gas application
Table 11-4: Accuracy (at calibration conditions):
RCCS30 0.008 g/cm³ ------ ------
RCCS31 0.004 g/cm³ 0.001 g/cm³ ------
RCCS32 0.004 g/cm³ 0.001 g/cm³ 0.0005 g/cm³
RCCS33 0.004 g/cm³ 0.001 g/cm³ 0.0005 g/cm³
RCCx34 0.003 g/cm³ 0.001 g/cm³ 0.0005 g/cm³
RCCx36 0.0022 g/cm³ 0.001 g/cm³ 0.0005 g/cm³
RCCx38 0.0015 g/cm³ 0.001 g/cm³ 0.0005 g/cm³
RCCx39 0.0015 g/cm³ 0.001 g/cm³ 0.0005 g/cm³
RCCx39/IR 0.0015 g/cm³ ------ ------
RCCx39/XR 0.0015 g/cm³ ------ ------
Repeatability:
- RCCS32-33, RCCx34-39/XR : ± 0.0005 g/cm³ (Std, /K4)
Static pressure effect:
Compensated if static pressure is set in the menu
Specication of high performance density measurement
(option /K6):
Ambient temp. range : -10°C to 50°C
Fluid temp. range : -50°C to 150°C
Minimum ow rate for specied accuracy:
- RCCx36 to RCCx39 : 700 kg/h
- RCCx34 : 140 kg/h
- RCCS33 : 90 kg/h
- RCCS32 : 37 kg/h
Maximum ow rate : Qnom
Repeatability : ± 0.0002 g/cm³
Temperature measurement: ±0.5°C ±0.2% of reading
Density accuracy : ±0.0005g/cm³ (liquids not aerated /
no gas in the liquid)
Process temperature inuence :
0.000015 g/cm³ * abs(Tuid-20°C)
-1.0
-0.9
-0.8
-0.7
-0.6
-0.5
-0.4
-0.3
-0.2
-0.1
0
0.1
0.2
0.3
0.4
0.5
0.6
0.7
0.8
0.9
1.0
0 50 100 150
Flow in % of Qnom
Error in %
F10.EPS