Reads the actual errors, alarms and warnings (EAW). In Event
Overview the number of the actual EAWs is shown. In
or the description of the actual events can be found.
Shows the latest 10 EAWs. By comparing this with
it can be checked if certain EAWs have vanished.
Select:
If is selected the latest 10 EAWs are cleared.
In the number of EAWs after the last clearance
can be viewed. In , or the description of the
events after the last clearance can be found.
Select:
If is selected the user is asked to ll in a date (mm/dd/yyyy).
If a date is set the errors and warnings are stored in the
with this date.
In the number of EAWs of the last clearance of the
can be viewed. In or the
description of the events can be found. The last 10 entries can be
selected by date.
Select:
If t is selected, all error and alarm status are checked.
If is selected, the existing errors and alarms are indicated in
succession.
These functions enable you to simulate the I/O signals (loop test).
The outputs can be checked with a multimeter.
Refer to the connecting diagrams in chapter 4.8.6. The test can be
cancelled by pressing any key.
Setting range: 2 to 22 mA
Setting range: 2 to 22 mA
Setting range: 0 to 10000 Hz
Setting range: On Active or Off Active
Setting range: 0 to 2000 Hz
Setting range: On Active or Off Active
Indication range: Open or Short