The following table shows possible countermeasures :
Alarm (A);
Warning
(W)
Error (E)
Name Meaning Description Countermeasure
A-01 MF 1 criteria Mass ow L Alarm Mass ow below MF 1 criteria Increase mass ow / Change setting
A-02 MF 2 criteria Mass ow H Alarm Mass ow above MF 2 criteria Reduce mass ow / Change setting
A-03 VF 1 criteria Volume ow L Alarm Volume ow below VF 1 criteria Increase volume ow / Change setting
A-04 VF 2 criteria Volume ow H Alarm Volume ow above VF 2 criteria Reduce volume ow / Change setting
A-05 Dens. 1 criteria Density L Alarm Density below Dens. 1 criteria Increase density / Change setting
A-06 Dens. 2 criteria Density H Alarm Density above Dens. 2 criteria Reduce density / Change setting
A-07 Temp. 1 criteria Temperature L Alarm Temperature below Temp. 1 criteria Increase temperature / Change setting
A-08 Temp. 2 criteria Temperature H Alarm Temperature above Temp. 2 criteria Reduce temperature / Change setting
A-09 Conc. 1 criteria Concentration L Alarm Concentration below Conc. 1 criteria Increase concentration / Change setting
A-10 Conc. 2 criteria Concentration H Alarm Concentration above Conc. 2 criteria Reduce concentration / Change setting
A-11 Netow 1 crit. Net ow L Alarm Net ow below Net ow 1 crit. Increase net ow / Change setting
A-12 Netow 2 crit. Net ow H Alarm Net ow above Net ow 2 crit. Reduce net ow / Change setting
A-13 Total thr exceed Total threshold exceed
Totalizer value exceeds total
threshold
Reset totalizer / Change setting
A-14 Slug detection Slug ow detected Slug criterion is reached Reduce gas bubbles in medium
A-15
Empty pipe
det.
Empty pipe detectd Empty pipe criterion is reached Fill pipe
A-16 Corrosion det. Corrosion detected Corrosion criterion is reached Change detector
W-01 Density low Density lower 0.3 kg/l Density lower than 0.3 kg/l Fill the detector with uid
W-02 Fixed dens act. Fixed Dens selected Parameter Dens x val sel is enabled
Set parameter Dens x val sel to
inhibit
W-03 Fixed temp act. Fixed Temp selected Parameter Temp x val sel is enabled
Set parameter Temp x val sel to
inhibit
W-04 Fixed MF active Fixed MF selected Parameter MF x val sel is enabled Set parameter MF x val sel to inhibit
W-05 Autozero running Autozero is running …
Autozero process is performed,
started by user action or status input
Wait until process stops
(30sec or 180sec max)
W-06 AZ val error
Autozero value out of
range
Auto zero value is out of the internal
detector depending range
Stop ow during auto zero.
Check detector installation
W-07 AZ uct error
Autozero uctuation out
of range
Auto zero uctuation is out of the
internal detector depending range
Stop ow during auto zero. Check
electrical installation, vibrations and
density.
W-08 Sensor simul. Sensor simulation active
Fixed frequency or xed phase dif-
ference is enabled
Set Sensor simulation to inhibit
W-09 Fixed Prim Out Analog output 1 xed
Analog output 1 xed at 4mA for
multidrop operation
Set HART polling address to 0
(Multidrop operation stops)
W-10 Prim Out sat. Analog output 1 saturated
Analog output 1 reaches less than
3.8mA or more than 20.5mA
Increase or decrease current level of
output 1
W-11 Sec Out sat. Analog output 2 saturated
Analog output 2 reaches less than
3.8mA or more than 20.5mA
Increase or decrease current level of
output 2
W-12 Tert Out sat.
Pulse/Status output 1
saturated
Relation between pulse width and
pulse rate of Pulse/Status output 1
is too high
Decrease length of pulse width or
increase amount of pulse rate of pulse
output 1
W-13 Quart Out sat.
Pulse/Status output 2
saturated
Relation between pulse width and
pulse rate of Pulse/Status output 2
is too high
Decrease length of pulse width or
increase amount of pulse rate of pulse
output 2
E-01 Frequency fault Frequency fault detected Driving frequency out of range Check cable (Remote) and detector
E-02 Signal fault Signal fault detected Phase difference out of range Check cable (Remote) and detector
E-03 EEPROM fault EEPROM fault detected CRC-Error Contact Yokogawa service
E-04 CPU fault CPU fault detected Micro processor is not working Contact Yokogawa service
E-05 DSP fault DSP fault detected Micro processor is not working Contact Yokogawa service
E-06 Sensor 1 fault
Sensor 1 Signal fault
detected
Sensor 1 signal below 7% for 3 min Check cable (Remote) and detector
E-07 Sensor 2 fault
Sensor 2 Signal fault
detected
Sensor 2 signal below 7% for 3 min Check cable (Remote) and detector
E-08 Temp sens fault
Temperature Sensor
Fault detected
T<-220°C or T>550°C
Check cable (Remote) and detector
Check temperature