 The highest measured medium temperature can be read in this
parameter.
This parameter can not be reset or changed by the user.

This option is factory set by order. If this function is selected nd the following restrictions:
- Density measurement is not available; the reference density value is always used
and displayed.
- Normal volume ow:
- Set reference density to normal density.
- Select normal volume (ow) units for volume, total volume, pulse output.
- Standard volume ow:
- Set reference density to standard density.
-
Select standard volume (ow) units for volume, total volume, pulse output.
- Volume ow:
- Set reference density to actual density.
- Select volume (ow) units for volume, total volume, pulse output.
- Corrosion detection is not available.
- Empty pipe detection is not available.
- Slug detection is not available
The instrument is mass balanced with air at normal pressure. It can be used for liquids with reduced accuracy.
The minimum URV-LRV range for mass ow, volume ow and net ow is 1% of Qnom.
If Heat of Combustion unit is changed between mass related units (MJ/kg, Btu/lb) and volume
related units
(MJ/m³(N), Btu/Scuft), there is no conversion of Heat of Combustion value.
In this case, Heat of Combustion has to be set manually to the correct value.