To set the functions of net ow measurement.
Net ow is only available if concentration measurement is activated
(option /Cxx).
(only available for option /CST)
Select:
When is set to :
can be selected
When is set to :
can be selected
Select:
(low range value)
Set a net ow value to the 4 mA or 0 Hz.
(upper range value)
Set a net ow value to the 20 mA or xxxxx = Hz.
Abs(URV-LRV) ≥ 5% of Qnom.
Set the damping time in s. Damping time effects outputs and display.
After 5 times damping time the given value is displayed (t=63%).
Set the low cut effects output and display. A hysteresis of 0.05% of
Qnom is given by the system.
Select:
Select a net ow alarm level. What happens in case of an alarm has
to be dened in Status out (7.12, 7.13)
Select:
Restrictions:
All alarms have to be in 0-110% of
URV-LRV
Alarm criteria 2 must always be
larger than alarm criteria 1.
Select a net ow alarm level. What happens in case of an alarm has
to be dened in Status out (7.12, 7.13)