(only for option/CST or /Cxx)
To set the functions of concentration measurement.
For xx=00 to 99 a customer specic or dened table is stored in the
converter in factory.
For /CST temperature coefcients have to be determined by the customer.
(only available for option/CST)
Select:
Select:
(low
range value)
Set a concentration value to the 4 mA or 0 Hz
(upper range value)
Set a to the 20 mA or xxxxx = Hz.
Set the damping time in s. Damping time affects outputs and display.
After 5 times damping time the given value is displayed (t=63%).
Set the low cut. Low cut affects outputs and display.
Select:
Select a concentration alarm level. What happens in case of an alarm
has to be dened in Status out (7.12, 7.13).
Select:
Restrictions:
All alarms have to be in 0-110%
of URV-LRV
Alarm criteria 2 must always be
larger than alarm criteria 1.
Select a concentration alarm level. What happens in case of an alarm
has to be dened in Status out (7.12, 7.13).
(only for option /CST)
To setup the coefcients used for calculation of concentration using the
Standard Concentration Measurement.
Set the temperature where the reference density of both components of
the mixture has been determined.
Set the density of the carrier liquid determined at the reference temperature.
Set the linear temperature coefcient of the density of the carrier liquid.
Set the squared temperature coefcient of the density of the carrier liquid.
Set the density of the product determined at the reference temperature.
Set the linear temperature coefcient of the density of the product.
Set the squared temperature coefcient of the density of the product.
For more information on concentration measurement see TI "Concentration Measurement" TI 01R04B04-04E-E.