Instruction Manual
7-93
IM 01R04B04-00E-E 8th edition March 01, 2011 -00
All Rights Reserved. Copyright © 2003, Rota Yokogawa
To set the pulse/status output 2.
Select:
If Pulse is selected, Status out 2 (SO) is not available.
If is selected, pulse parameters are not available.
Select:
Numbers may be displayed if selected via HART.
If or are selected, pulse
output only works at forward ow.
If or are selected, pulse
output only works at reverse ow.
To select the pulse output mass
or volume has to be selected.
If mass ow or volume ow is
selected the output delivers only
frequency.
Select:
If is
or , only can be selected.
If is
or only can be selected.
only for gas application.
only for gas application or if Reference density
value is selected.
The Pulse rate can be set according to the unit set in
Maximum pulse rate is 2000 Hz.
Select:
The output levels depending on this parameter are shown on page 7-91.
Select:
Select:
If is selected, in case of alarm the pulse output is stopped.
If is selected, the pulse output keeps on measuring.
If is selected, the pulse output keeps the value before entering the
alarm/error state.
NOTE