(using the display keys, equivalent via HART)

 or 


Select : 

 or 


Select : 

 → 
 → 
 → 
 → 
 → 
 → 
 → 
 → 
 → 
 → 
 → 
→ 
Notice : By selecting a variable (mass or volume)
which can be totalized automatically the pulse
output is selected.
 Select:






If  is selected, status output becomes active for forward
ow and not active for reverse ow, if parameter  set to
. It is the other way round, if parameter  is set to
 (see 7.17).
If  is selected, status output becomes active, when
the totalizer reaches the value set to parameter  (see
7.17).
If an alarm function is selected, status output becomes active, if the
concerning alarm occurs.
If  is selected, status output becomes active, if at least one
alarm occurs. The same applies for errors with the setting All errors.
For  the status output becomes active, if at least
one alarm or one error occurs.
 Select: 
The output levels depending on this parameter are shown on page 7-91.
If  is selected, under  the
 and

have to be dened.
Under ´Detailed Setup´


 and

have to be dened.
If the pulse output is set under, the
pulse width is the width set under 
If the pulse width is not congured under 
 menu the  will be set default to
1ms.
If ´frequency´ is selected the frequency generated
at URV (upper range value) has to be dened.
Default is URV = 1000Hz.


For pulse output the mass (or volume) is internally
totalized and if the increment is reached, one
pulse is sent.
For a frequency output the measured value is
calculated in actual frequency according to the
denition of the user (e.g. 10000 Hz = 54321 kg/h)
with 50% duty cycle.
Set as frequency output the pulse length is changing
with the change of the measured value.