 To set the current output 2.
 Select: 

Numbers may be displayed if selected via HART.
If is selected, output is 4 mA.

(Detailed setup, congure out-
put/input)
If is selected (meter without option /NA):

If is selected (meter with option /NA):


(low limit)
(Detailed setup, congure output/
input)
The alarm is active if the
measured value is lower than the
low limit.
If is selected (meter without option /NA):
Measuring range from 3.8 - 20.5 mA
Alarm level: < 3.6 mA or > 21.0 mA
If is selected (meter with option /NA):
Measuring range from 2.7 - 21.3 mA
Alarm level: < 2.4 mA or > 21.6 mA

(high limit)
(Detailed setup, Congure output/
input)
The alarm is active if the
measured value is higher than
the high limit.
If is selected (meter without option /NA):
Measuring range from 3.8 - 20.5 mA
Alarm level: < 3.6 mA or > 21.0 mA
If is selected (meter with option /NA)
Measuring range from 2.7 - 21.3 mA
Alarm level: < 2.4 mA or > 21.6 mA