 To set the functions of temperature measurement.
Notice: Standard range: -200 to 230°C
High temp. range (/HT): 0 to 400°C
 Select: 
 Select: 

(low range value)
Set a temperature value to the 4 mA or 0 Hz

(upper range value)
Set a temperature value to the 20 mA or xxxxx Hz.
Abs(URV-LRV) 10K
 Set the  in s. Damping time affects outputs and display.
After 5 times damping time the given value is displayed (t=63%).
 Select: 
 Select a  level.
What happens in case of an alarm has to be dened in Status out (7.12,
7.13).
 Select: 

Restrictions:
All alarms have to be in 0-110%
of URV-LRV Alarm criteria 2
must always be larger than alarm
criteria 1.
Select a  level.
What happens in case of an alarm has to be dened in Status out (7.12,
7.13).
 Select: 
 Select a temperature value in case of malfunction of the built in
temperature sensor.
 The ROTAMASS temperature sensor is installed outside of the measuring
tubes. So at high or low uid temperature the temperature measurement
may show a deviation as a function of the quality of heat insulation. This
function allows correcting temperature measurement for better accuracy
in mass ow and density.
Note: It is not recommended to use this function unless correct
measurement over the complete process temperature range can be
conrmed. It is not sufcient to proof correct measurement at one
temperature or in a very narrow temperature range unless the process
operates exclusively within this range.

 To set the function of the medium velocity.
Not available for gas measurement (/GA)
 Select: 