Instruction Manual


7-84
IM 01R04B04-00E-E 8th edition March 01, 2011 -00
All Rights Reserved. Copyright © 2003, Rota Yokogawa

 To set the functions of volume ow measurement. Notice: Volume ow
is derived density and mass ow measurement.
Errors in density measurement (e.g. from gas bubbles) effects the
accuracy of volume ow. For liquid it is recommended to use mass ow
as original measurement. Density below roh = 0.3 kg/l volume ow is
set to zero (for liquids)
 Select:







only for gas application.

 only for gas application or if Reference density
value is selected.
 Select; 

(low range value)
Set a volume ow value to the 4 mA or 0 Hz. LRV can also be negative
to measure reverse ow.

(upper range value)
Set a volume ow value to the 20 mA or xxxxx Hz.
Abs(URV-LRV) 5% of Qvnom for liquids.
For gas application Abs(URV-LRV) 1%
 Set the damping time in s. Damping time affects outputs and display.
After 5 times damping time the given value is displayed (t=63%)
 Set the low cut. Low cut affects outputs and display. Any low cut on
mass ow will affect volume ow also. A hysteresis of 0.05% of Qnom
on mass ow is given by the system.
 Select: 
 Select a volume ow alarm level. What happens in case of an alarm
has to be dened in Status (7.12, 7.13).
 Select: 

Restrictions:
All alarms have to be in 0-110%
of URV-LRV.
Alarm criteria 2 must always be
larger than alarm criteria 1.
Select a volume ow alarm level. What happens in case of an alarm
has to be dened in Status out (7.12, 7.13).