User guide
12-19
IM 04L41B01-01E
Specifications
1
2
3
4
5
6
7
8
9
10
11
12
App
Index
12.6 General Specifications
Construction
Item Specifications
Mounting Flush panel mounting (on a vertical plane) (excluding the desktop type)
Mounting angle Inclined backward up to 30 degrees from a horizontal plane.
Allowable panel thickness 2 to 26 mm
Material Case: Metal plate
Bezel and display cover: Polycarbonate
Color Case: Grayish blue green (Munsell 2.0B5.0/1.7 or equivalent)
Bezel: Charcoal gray light (Munsell 10B3.6/0.3 or equivalent)
Front panel Water and dust proof: Complies with IEC529-IP65 and NEMA No.250 TYPE 4 (except External Icing
Test) (Style number 2 or later), except side-by-side mounting
External dimensions 144 (W) × 144 (H) × 229 (D) mm (D: depth from the panel mounting plane)
Weight DX1000: Approx. 2.9 kg excluding options, DX1000N: Approx. 3.7 kg excluding options
Normal Operating Conditions
Item Specifications
Supply voltage 90 to 132, 180 to 250 VAC
Power supply frequency 50 Hz ± 2%, 60 Hz ± 2%
Ambient temperature 0 to 50°C
Ambient humidity 20 to 80%RH (at 5 to 40°C), 10 to 50% (at 40 to 50°C)
Vibration 10 to 60 Hz, 0.2 m/s
2
Shock Not allowed
Magnetic field 400 A/m or less (DC and 50/60 Hz)
Noise Normal mode (50/60 Hz)
DC voltage
The peak value including the signal must be less than 1.2 times the measuring range.
Thermocouple
The peak value including the signal must be less than 1.2 times the measuring thermal
electromotive force.
RTD 50 mV or less
Common mode noise 250 VACrms or less for all ranges (50/60 Hz)
Maximum noise voltage between channels
250 VACrms (50/60 Hz) or less
Mounting position Can be inclined up to 30 degrees backward. Left and right horizontal.
W
arm-up time At least 30 minutes after power on
Installation location Indoors
Operating altitude 2000 m or less
Power Supply
Item Specifications
Rated supply voltage 100 to 240 VAC
Allowable power supply voltage range
90 to 132, 180 to 264 VAC
Rated power supply frequency
50 Hz, 60 Hz
Power consumption Supply voltage LCD backlight off Normal
Maximum
100 VAC 15 VA 24 VA 45 VA
240 VAC 25 VA 32 VA 60 VA
Allowable interruption time Less than 1 cycle of the power supply frequency
Isolation
Item Specifications
Insulation resistance Between the Ethernet, RS-422/485, and insulation terminals and earth: 20 MΩ or greater at 500 VDC
Withstand voltage Between the power terminal and earth: 2300 VAC at 50/60 Hz for one minute
Between the contact output terminal and earth: 1600 VAC at 50/60 Hz for one minute
Between the measurement input terminal and earth: 1500 VAC at 50/60 Hz for one minute
Mutually between measurement input terminals: 1000 VAC (50/60 Hz) for one minute (excluding
the RTD input terminal of DX1006, DX1012,
DX1006N and DX1012N)
Between the remote input terminal and earth: 1000 VDC for one minute
Between the pulse input terminal and earth: 1000 VDC for one minute
Ground Grounding resistance: 100 Ω or less










