3.13 Managing the Input Calibration Interval (/CC1 option; release numbers 4 and later)
Procedure
Viewing the Calibration Notification Screen
The calibration notication screen only appears over operation screens. It will continue
to appear until you perform calibration (Finish Calibration).
You can close the calibration notication screen by pressing ESC or DISP/ENTER
.
Operation Example
When Notication is set to 1 (DayBfr) and Renotication is set to 8h
Notification screen
display start
Calibration due date
Notification screen
appears
Renotification Renotification Renotification Renotification
00:00 08:00 16:00 00:00 08:00
Notification (DayBfr) setting
Notification screen
appears
Notification screen
appears
Notification screen
appears
Note
The calibration notification screen will close if you change screens while it is displayed.
If the calibration due date passes before you complete calibration, that information will be
recorded in the login log (the operation log on DXs with the /AS1 option).