User guide

Measurement Channels and Alarms
IM 04L41B01-01E
3-7
1
2
3
4
5
6
7
8
9
10
11
12
App
Index
3.5 Setting the Auxiliary Alarm Function
Setup Screen
Press MENU (to switch to setting mode), hold down FUNC for 3 s (to switch to basic
setting mode), and select the Menu tab > Alarm > Basic settings; Switch, Relay; or
Hysteresis.
• Basic settings
• Switch, Relay
• Hystersis
Setup Items
Basic settings > Reflash
To set the reash operation on the alarm output relay, select On. The reash function
is set on the rst three output relays.
Setting Description
Off Reflash is not used.
On Reflash is used. The relays are deactivated for approximately 500 ms.
On-1s Reflash is used. The relays are deactivated for approximately 1 s.
On-2s Reflash is used. The relays are deactivated for approximately 2 s.
Basic settings > Rate of change
Decrease
Set the interval for the rate-of-change calculation of the low limit on rate-of-change
alarm in terms of the number of sampled data points (1 to 32). The actual interval is
obtained by multiplying the value specied here by the scan interval.
Increase
Set the interval for the rate-of-change calculation of the high limit on rate-of-change
alarm in the same manner as the interval for the low limit on rate-of-change alarm.