Manual

4-4
SM 04L41B01-01E
7.
Press the End #4 soft key.
DX1000/DX1000N DX2000
Saving Adjusted Values
8.
Use the left or right arrow key to choose Yes (save calibration values) or No (do
not save), then press the DISP/ENTER key.
DX1000/DX1000N DX2000
9.
Repeat steps
2
8
to adjust all A/D No.
When finished, press the ESC key.
Concluding Adjustments
10.
Turn OFF the power switch.
Note
After adjusting the DX1000N, turn ON the power switch then close the bezel. If the power
switch is OFF when you close the bezel, you will have to open it again the next time you use
the instrument.
4.2 Adjusting the A/D Converters