Owner manual

6-11
IM 04L41B01-17E
Specifications
1
2
3
4
5
6
App
Index
Extended Hold Registers (Shared with the Modbus slave function; release
number 3 or later)
The following hold registers have been added. You can perform a portion of the
operations by writing in the registers. You can write to these registers when a DX with the
/AS1 advanced security option is in operation mode.
Internal switch
L
ot number or lot number for each batch group (when the multi batch function (/BT2
option) is in use)
Batch number or batch group number for each batch group (when the multi batch
function (/BT2 option) is in use)
Recording (memory sampling) start and stop or recording (memory sampling) start
and stop for each batch group (when the multi batch function (/BT2 option) is in
use)
Alarm ACK
Alarm display reset
Computation start, computation stop, computation reset, computation dropout ACK,
and computation reset for each batch group (when the multi batch function (/BT2
option) is in use)
Manual sampling, event data sampling start trigger, and snapshot
Message and free message writing or message and free message writing for each
batch group (when the multi batch function (/BT2 option) is in use)
Event edge switch
Event level switch
6.3 Modbus Protocol Specifications