Owner manual
4-7
IM 04L41B01-17E
Responses
1
2
3
4
5
6
App
Index
tt
Command name (
SR
,
SA
···,
XA
,
XI
···)
sss···s
Setting/basic setting data (variable length, one line)
• Example
EA
SR001,VOLT,20mV,0,20
SR002,VOLT,20mV,0,20
···················
EN
Decimal Point Position/Unit Information
• The FE command is used to output the data.
• You can use the CB command to specify whether to output the data of measurement
channels set to skip and computation channels set to OFF.
•
Syntax
The data is output for each channel in the following syntax.
EACRLF
s_cccuuuuuu,ppCRLF
··················
ENCRLF
s
Data status (
N
,
D
, or
S
)
N:
Normal
D:
Differential input
S:
Skip (When the measurement range is set to SKIP for a
measurement channel or when the channel is turned OFF for a
computation channel)
ccc
Channel number (3 digits)
001
to
048:
Measurement channel
101
to
160:
Computation channel
201
to
440:
External input channel
uuuuuu
Unit information (6 characters, left-justified)
mV____:
mV
V_____:
V
^C____:
°C
xxxxxx:
(User-defined character string)
pp Decimal point position (
00
to
04
)
No decimal (00000) for 00.
One digit to the right of the decimal (0000.0) for 01.
Two digits to the right of the decimal (000.00) for 02.
Three digits to the right of the decimal (00.000) for 03.
Four digits to the right of the decimal (0.0000) for 04.
_
Space
•
Example
EA
N 001mV ,01
N 002mV ,01
EN
4.2 Output Format of ASCII Data