Owner manual
3-27
Commands
IM 04L41B01-17E
1
2
3
4
5
6
App
Index
3.4 Setting Commands
Query
TT?
Example
Display waveform horizontally
, set the message
direction to vertical, and display waveforms by
clearing the current waveforms at memory start.
TTHORIZONTAL,ON,VERTICAL
Description
When using the /BT2 multi batch option, p2 is
fixed at ON.
SE Sets the line width and the
number of grids to use on the
trend graph
Syntax
SE p1,p2<terminator>
p1
T
rend line width (1 to 3) [dots]
p2
Number of grids (4 to 12, AUT
O)
Query
SE?
Example Set the trend waveform line width to 1 dot and
the number of grids to 10.
SE1,10
TB Sets the bar graph display
Syntax
TB p1<terminator>
p1
Bar graph display direction
HORIZONTAL
VERTICAL
Query
TB?
Example
Display the bar graph horizontally
.
TBHORIZONTAL
SB Sets the bar graph for a channel
Syntax
SB p1,p2,p3<terminator>
p1
Measurement, computation, or external
input channel number
p2
Bar graph base position
NORMAL
Normal (lower limit)
CENTER
Center
LOWER
Lower limit
UPPER
Upper limit
p3
Number of scale divisions (4 to 12)
Query
SB[p1]?
Example
Set the number of scale divisions on the bar
graph for channel 002 to five, and display the bar
graph from the span lower limit (the scale lower
limit if scale is enabled).
SB002,NORMAL,5
Description You can specify computation channels on
models with the /M1 or /PM1 math option. You
can specify external input channels on models
with the /MC1 external input channel option.
TN Sets a scale
Syntax
TN p1,p2,p3<terminator>
p1
Measurement, computation, or external
input channel number
p2
Display position (OFF, 1 to 10)
p3
Number of divisions (4 to 12, C10)
Query
TN[p1]?
Example
Set the scale position for channel 003 to 2, and
the number of divisions to 10.
TN003,2,10
SV Sets a measurement channel’s
moving average
Syntax
SV p1,p2,p3<terminator>
p1
Measurement channel number
p2
Moving average (OFF
, ON)
p3
Number of moving average samples (2 to
400)
Query
SV[p1]?
Example
Set the number of moving average samples for
channel 002 to 12.
SV002,ON,12
SC Sets a channel display color
Syntax
SC p1,p2<terminator>
p1
Measurement, computation, or external
input channel number
p2
Display color (see SL
(sets a trip line))
Query
SC[p1]?
Example
Set the channel 002 display color to blue.
SC002,BLUE
Description You can specify computation channels on
models with the /M1 or /PM1 math option. You
can specify external input channels on models
with the /MC1 external input channel option.
TA Sets an alarm point mark
Syntax
TA p1,p2,p3,p4,p5,p6,p7<terminator>
p1
Measurement, computation, or external
input channel number
p2
Mark type
ALARM
Alarm mark
FIXED
Fixed mark
p3
Scale board (ON, OFF)
p4
Alarm level 1 color (AUTO or 24 colors (see
NL; sets a trip line))
AUTO
The same color as the alarm color
p5
Alarm level 2 color (AUT
O or 24 colors (see
NL; sets a trip line))
AUTO
The same color as the alarm color
p6
Alarm level 3 color (AUT
O or 24 colors (see
NL; sets a trip line))
AUTO
The same color as the alarm color
p7
Alarm level 4 color (AUT
O or 24 colors (see
NL; sets a trip line))
AUTO
The same color as the alarm color
Query
TA[p1]?
Example
Display alarm marks on the channel 004 scale.
TA004,ALARM,ON