For parameters p7, p8, and p9, either set all
three parameters or omit all three parameters.
When Setting Channels to Square Root
Computation
Syntax
SR p1,p2,p3,p4,p5,p6,p7,p8,p9,p10,
p11<terminator>
p1
Measurement channel number
p2
Setting type (SQR
T)
p3
Measurement range
p4
Span lower limit
p5
Span upper limit
p6
Scaling lower limit (-30000 to 30000)
p7
Scaling upper limit (-30000 to 30000)
p8
Scaling decimal place (0 to 4)
p9
Unit (up to 6 characters)
p10
Low-cut function (OFF, ON)
p11
Low-cut point (0 to 50)
Query
SR[p1]?
Example
Convert the DC voltage measured on channel
001 to an amount of flow using the square root
computation. Set the measurement range to 6 V,
the span lower limit to 1 V, the span upper limit to
5 V, the scaling lower limit to 10.0 m
3
/s, and the
scaling upper limit to 100.0 m
3
/s.
SR001,SQRT,6V,1000,5000,100,1000,1,
m3/s
Description
Y
ou cannot use this command while recording
(memory sampling) or computation is in
progress. When using the /BT2 multi batch
option, you cannot use this command if any
of the batch groups is recording (memory
sampling).
Set p1 and p3 by referring to the table in
section 3.3.
For parameters p4 and p5, enter values with
five digits or less excluding the decimal point.
For parameters p6, p7, and p8, either set all
three parameters or omit all three parameters.
For 1-5V DC Voltage Input
Syntax
SR p1,p2,p3,p4,p5,p6,p7,p8,p9,p10
<terminator>
p1
Measurement channel number
p2
Input type (1-5V)
p3
Measurement range (1-5V)
p4
Span lower limit (800 to 5200)
p5
Span upper limit (800 to 5200)
p6
Scaling lower limit (-30000 to 30000)
p7
Scaling upper limit (-30000 to 30000)
p8
Scaling decimal place (0 to 4)
p9
Unit (up to 6 characters)
p10
Low-cut function (ON, OFF)
Query
SR[p1]?
Example
Set the channel 005 input type to 1-5V
, the span
lower limit to 1 V
, the span upper limit to 5 V, and
turn the 1-5V low-cut function ON.
SR005,1-5V,1-5V,1000,5000,,,,,ON
Description You cannot use this command while recording
(memory sampling) or computation is in
progress. When using the /BT2 multi batch
option, you cannot use this command if any
of the batch groups is recording (memory
sampling).
Set p1 by referring to the table in section 3.3.
For parameters p4 and p5, enter values with
four digits or less excluding the decimal point.
For parameters p6, p7, and p8, either set all
three parameters or omit all three parameters.
SO Sets a computing equation
Syntax
SO p1,p2,p3,p4,p5,p6,p7<terminator>
p1
Computation channel number
p2
Computation (ON, OFF)
p3
Computing equation (up to 120 characters)
p4
Span lower limit (-9999999 to 99999999)
p5
Span upper limit (-9999999 to 99999999)
p6
Span decimal place (0 to 4)
p7
Unit (up to 6 characters)
Query
SO[p1]?
Example
Compute the sum of channels 001 and 002
using channel 106. Set the span lower limit to
-10.0000, the span upper limit to 15.0000, and
the unit to V
.
SO106,ON,001+002,-100000,150000,4,V
Description
Y
ou can use this command on models with the
/M1 or /PM1 math option.
Y
ou cannot use this command while recording
(memory sampling) or computation is in
progress. When using the /BT2 multi batch
option, you cannot use this command if any
of the batch groups is recording (memory
sampling).
For details on computing equations, see the
DX1000/DX1000N or DX2000 User’s Manual.
Set p1 by referring to the table in section 3.3.
For parameters p4 and p5, enter values with
seven digits or less, excluding the decimal, for
negative numbers and with eight digits or less
for positive numbers.
For parameters p4, p5, and p6, either set all
three parameters or omit all three parameters.
ER Sets the range of an external
input channel
Syntax
ER p1,p2,p3,p4,p5,p6<terminator>
p1
External input channel number
p2
External input channel (ON, OFF)
p3
Span lower limit (-30000 to 30000)
p4
Span upper limit (-30000 to 30000)
p5
Decimal place (0 to 4)
p6
Unit (up to 6 characters)
Query
ER[p1]?
3.4 Setting Commands