For the manual adjustment procedure according to the two cases described the following remarks have to be
taken into account:
After manual adjustment the ow-scale of the indicator is no longer valid.
The display will only indicate in % of the new ow range.
Switching of units is impossible.
The indicator can be always resetted to the original adjustment according to manufacturers
calibration at any time.
The given procedures will only adjust the current output and the display to the new measuring
range for a different uid and/or new process conditions.
The result of this adjustment is NOT a calibration! If proof of the new adjustment is requested a
real calibration by comparison to a standard has to be made after adjustment!
The following interactions with other functions apply:
Interaction with other functions:
Action Function Effect
1. Setting of manual adjustment values F52 -5... Manual adjustment table is overwritten
2. Activating of manual adjustment table F51 -2 - Manual adjustment table active
- %-indication only
- Totalizer does not count
- No other ow rate unit selectable
- Function F64 for option /A2 has no effect,
with manual adjustment
3.
Deactivating of manual adjustment
table
F51 -1
- Standard adjustment table active
- F11 is set to ow rate
- Flow unit is the same as before activating
manual adjustment
- Totalizer value is the same as before
activating manual adjustment
Due to the limitations described it is strongly recommended to order a new scale and EEPROM from
manufacturer (this gives a real new ow scale without new calibration) or to order a new calibration by
the manufacturer together with a new scale and EEPROM for the new uid and/or new process conditions
(this gives new adjustment plus new calibration).