Remote control function (/R1 option)
Item Description Reference 
Section
Remote control function (/R1 option)
• Assign functions to the remote control input terminals 7.18
  Use Remote in Basic Setting mode to set the function to be assigned to the remote control input 
terminal.
•
 Secondary chart speed 6.9
  If “chart speed switching” is assigned, use Aux >
 Chart 2 in Setting mode to set the secondary 
chart speed.
• Switching between start printout and start printout 2 and between end printout and end printout 
2 (/BT1 option)
  If BatchCMT switch is assigned
  Use Batch
 > Dual comment
 in Basic Setting mode to enable or disable start printout 2 and end 
printout 2.
7.24
  If Use is selected, set Batch >
 Detail > Start2 and End2 in Setting mode. 6.14
  If Not is selected, the Batch >
 Detail > Start2 and End2 items in Setting mode do not appear.
Calibration Correction (/CC1 Option)
Item Description Reference 
Section
Calibration correction function
Use Personal
ize
 > Add function > Calibration in Basic Setting mode and select Use or Not. 7.15
If Use is selected,
• Use Cal
ibration in Basic Setting mode to set the correction mode and the number of calibration 
points.
7.23
• Use Cal
ibration in Setting mode to set the measured value and correction value for each 
channel.
6.13
If Not is selected, the Cal
ibrat
ion item does not appear.
Header Printout (/BT1 Option)
Item Description Reference 
Section
Setting start printout and end printout
Use Batch in Basic Setting mode to set whether to use the start printout and end printout. If Use is 
selected, select a Lot No. from 4 digits, 6 digits, or OFF.
7.24
If Use is selected,
Use Batch
 >
 Batch No. in Setting mode to set the batch number. 6.14
Use Batch >
 Lot No. in Setting mode to set the lot number.
Use Batch >
 Detail > Start, End in Setting mode.
If Not is selected, the Batch >
 Lot No. item in Basic Setting mode and the Batch item in Setting 
mode do not appear.
Switch between start printout and start printout 2 and between end printout and end printout 2 (/R1 option)
• Switch settings
  Assign “BatchCMT switch” to the remote control input terminal.
7.18
  Use 
Batch >
 Dual comment in Basic Setting mode to set whether to use start printout 2 and end 
printout 2.
7.24
  If Use is selected, set Batch >
 Detail > Start2 and End2 in Setting mode. 6.14
  If Not is selected, the Batch >
 Detail > Start2 and End2 items in Setting mode do not appear.
• Executing the switch 1.4
  The switch occurs according to the status of the “BatchCMT switch” signal when “Record On/Of
f” 
or “Priority R_RCD” signal assigned to the remote control input terminal rises or falls.
Printout of messages including measured/computed values
•
 Setting message strings 6.8
  Use Aux >
 Message in Setting mode to enter the message to print out.
• Setting the message format
  Use Batch
 > MSG format
 in Basic Setting mode and select whether to use the message format. 7.24
  If Use is selected, use Batch >
 MSG format > Message No. in Setting mode to set the message 
format.
6.15
  If Not is selected, the Batch >
 MSG format item does not appear.
• Executing message printout 3.10
  Use Func >
 Message in Operation mode to execute the message printout.
1.10 Function Setup Guide