User Manual
1-31
IM 04P01B01-01E 
Functional Explanation and Setup Guide
For the procedure to set the functions, see section 1.10, “Function Setup Guide.”
1
2
3
4
5
6
7
8
9
10
11
12
App
Index
•  Computation Start/Stop (/M1 Option)
•  Remote input signal: Rising edge signies start; falling edge signies stop
•  Starts/stops the computation.
•  If the computation is started, applying a rising edge signal produces no effect. If the 
computation is stopped, applying a falling edge signal produces no effect.
•  Computation Reset (/M1 Option)
•  Remote input signal: Trigger
•  The computed results of computation channels are reset. This is valid only while 
the computation is stopped. For all other cases, applying the remote signal 
produces no effect.
•  Priority to Remote Recording (/BT1 Option)
•  Remote input signal: Edge (rising/start or falling/stop)
•  Starts/stops recording.
•  When started with a remote signal (on a remote signal rise), stop per key operation 
or communication is disabled.
•  Switching Batch Comment (/BT1 Option)
•  Remote input signal: Level
•  Switches between Start printout and Start printout 2, and between End printout 
and End printout 2 depending on the status of the batch comment switching signal 
when recording is started/stopped remotely.
•  When starting/stopping by key operation, performs Start printout and End printout.
Remote Signal (Edge, Trigger, and Level)
The above actions are carried out on the rising or falling edge of the remote signal (edge), 
the ON signal lasting at least 250 ms (trigger), or the ON/OFF signal (level).
Rising/Falling edge Trigger Level
Rising Falling
250 ms or more
Operates at the 
secondary chart speed
250 ms or more
For contact inputs, the remote signal rises when the contact switches from open to closed 
and falls when the contact switches from closed to open. For open collector signals, the 
remote signal rises when the collector signal (voltage level of the remote control terminal) 
goes from high to low and falls when the collector signal goes low to high.
<Related
 T
opics> Assigning functions to the remote control input terminals: Section 7.18 
Setting 
the secondary chart speed: Section 6.9
1.8 Remote Control Function (/R1 Option)