For the procedure to set the functions, see section 1.10, “Function Setup Guide.”
•   The Number of Channels Being Trend Recorded and the Fastest 
Recording Interval
  The fastest recording interval varies depending on the number of channels being 
trend recorded. Only display or printout is possible on measurement channels or 
computation channels that are not trend recorded.
Number of Channels Being Trend Recorded Fastest Recording Interval
6 channels or less 10 s
7 to 12 channels 15 s
13 to 18 channels 20 s
<Related Topics> Changing the channel recording color: Section 9.13 
Enabling/Disabling trend recording for each channel: Section 9.9
Printout
Various printouts can be executed in the same fashion as measurement channels.
The computed values of computation channels can be printed at constant intervals by 
using the periodic printout timer or the two TLOG timers. The TLOG timers can be used 
only on the computation channels. For the printout contents when printing using the 
TLOG timer, see appendix 1.
<Related
 T
opics> Setting the periodic printout: Sections 7.8, 9.6, and 9.15 
Setting the 
printout using the TLOG timer: Sections 9.6 and 9.12
Alarms
Up to 4 alarms can be assigned to each computation channel.
•  The alarm types are high limit alarm (H), low limit alarm (L), delay high limit alarm (T), 
and delay low limit alarm (t).
•  Hysteresis can be specied to the values for activating and releasing the alarm. 
The 
hysteresis applies only to high limit alarm (H) and low limit alarm (L). The hysteresis 
width can be set in the range of 0.0% (Off) to 1.0% of the recording span in 0.1 
steps. The setting applies to all high limit alarms and low limit alarms. By default, the 
hysteresis width is set to 0% (Off).
<Related
 T
opics> Setting alarms: Section 9.5 
Setting the alarm delay function: Sections 7.15 and 9.11 
Setti
ng the alarm hysteresis: Section 7.1
Display
Computation channels can be displayed on the VFD as with the measurement channels.
<Related Topics> Setting the bar graph display mode: Section 9.16
Starting/Stopping Computation
Computation can be started or stopped using keys.
<Related Topics> Starting/Stopping computation: Section 9.1