7.
  Press the ESC key to finish the adjustment. The screen returns to the 
Basic=P_Adj screen.
Select Hysteresis, Zero, or Full.
Change the value to adjust the pen 
position.
The dot printing position moves.
ESC/?
Basic=P_Adj
P_Adj=Hysteresis
P_Adj
Setting complete
Hysteresis=-2
Applying the Changes and Returning to Operation Mode
Press the   key or SHIFT +   key to select End and then press the   key. Press 
the 
 key to select Store and then press the   key. The changes are applied, and 
the screen returns to Operation mode. If you select Abort and press the 
 key, the 
changes are discarded, and the screen returns to Operation mode. Press the ESC key to 
return to the Basic=
 screen.
Note
When the Customized Menu is enabled, the operation is different.
•  When “P_Adj” in the Customized Menu is On
1. Perform steps 1-4 of “Adjusting the Dot Printing Position.”
2. When the instrument enters Basic Setting mode, “Menu kind” appears. Select “Free” 
then press a   key.
3. Perform the steps starting with step 5 of “Adjusting the Dot Printing Position.”
•  When “P_Adj” in the Customized Menu is Off
  See “Disabling the Customized Menu” or “Changing the Settings without Disabling the Basic 
Setting Mode Lock.”
<Related Topics> Enabling/Disabling the customized menu: Section 7.22 
Changing the settings without r
eleasing the basic setting mode lock: 
Section 7.22
11.6 Adjusting the Dot Printing Position (Dot Model)