User Manual
11-5
IM 04P01B01-01E 
Maintenance
1
2
3
4
5
6
7
8
9
10
11
12
App
Index
Temperature Measurement When Using an RTD
Decade resistance box
The resistance of three lead wires must be equal.
(Model 2793-01 from Yokogawa
 Meters & Instruments)
+/A
/b
Input terminals
–/B
Temperature Measurement When Using a Thermocouple
Thermocouple wires or TC extension wiresCopper wires
DC voltage standard
+
–
Input terminals
+ –
Thermocouple
wires
Copper
wires
(0°C standard temperature device ZC-114/ZA-10 by Coper Electronics)
RJC of TC input
As the measurement terminal of the recorder is generally at room temperature, the actual 
output of the thermocouple is different from the values given on the thermoelectromotive 
force table based on 0°C. The recorder performs compensation by measuring the 
temperature at the input terminal and adding the corresponding thermoelectromotive 
force to the actual output of the thermocouple. Therefore, when the measurement 
terminal is shorted (equivalent to the case when the detector tip is 0°C), the measured 
value indicates the temperature of the input terminal.
When calibrating the recorder, this compensation voltage (thermoelectromotive force of 
0°C reference corresponding to the input terminal temperature) must be subtracted from 
the output of the DC voltage standard before application. As shown in the figure, by using 
the 0°C standard temperature device to compensate the reference junction at 0 °C, you 
can input the thermoelectromotive force of 0 °C reference from the DC voltage standard 
and perform the calibration.
11.4 Calibrating the Recorder










