User Manual
IM 04P01B01-01E 
10-6
10.2 Troubleshooting Flow Charts
Is the input free of noise?
NO
YES
• Separate the input and 
 communication wiring from the 
 noise source.
• Ground the recorder properly.
• Use shielded cables for input and 
 communication wiring.
• Insert an isolation transformer 
 into the power line.
• Insert a power line filter into the 
 power line.
• If an inductive load is connected 
 to an alarm contact output, use a 
 surge suppresser on that line.
Contact your nearest 
YOKOGAWA dealer.
Display and other functions do not work properly.
Is the key lock released?
NO
YES
Release the key lock using the 
FUNC key (see section 3.13).
Contact your nearest 
YOKOGAWA dealer.
Keys do not work.
Is the customize 
menu released?
NO
YES
Disable the Customized Menu 
with a key operation (see section
7.22).
Contact your nearest 
YOKOGAWA dealer.
Some of the menus are not displayed.
Unable enter Basic Setting mode.