User Manual

IM 04P01B01-01E
10-2
10.1 A List of Error Messages
Code Message Explanation/Countermeasures
64 MATH upper and lower span values are equal. This is not allowed.
70 MATH constant description is incorrect. Incorrect computation constant syntax.
71 The range of the MATH constant is exceeded.
72 MATH channel is turned off
81 All space or ‘quit’ string cannot be specified.
86 The key-lock release password is incorrect. Enter the correct password.
87 This key is locked.
91 Password is incorrect. Enter the correct password.
100 IP address doesn’t belong to class A, B, or C.
101 The result of the masked IP address is all 0s or 1s.
102 SUBNET mask is incorrect. Set a correct subnet mask.
103 The net part of default gateway is not equal to
that of IP address.
Set the correct default gateway.
151 This action is invalid during calculation.
160 This action is invalid during chart end.
161 This action is invalid during pen hold.
162 Cannot set an number for a skipped data.
163 This action is invalid during record.
164 This action is invalid during manual printing.
165 This action is invalid during list printing.
166 This action is invalid during setup list printing.
167 This action is invalid during chart feed.
169 This action is invalid during ribbon hold.
170 This action is invalid during priority remote record. Recording stop via communications or key operation cannot be
accepted under the current settings. Perform the Stop using the
remote control function.
171 This action is invalid during batch. Cannot enter Basic Setting mode during header printout (until
End printout is finished).
Operation Errors
Code Message Explanation/Countermeasures
232 There is no available data. There is no data for periodic printout or data for calculating
TLOG when the timer expired.
Communication Errors
Code Message Explanation/Countermeasures
300 Command is too long.
301 Too many number of commands delimited with ’;’. Keep the number of commands separated by sub delimiters
under 10.
302 This command has not been defined.
303 Data request command can not be enumerated with
sub-delimiter.
350 Command is not permitted to the current user level.
351 This command cannot be specified in the current
mode.
352 The option is not installed.
353 This command cannot be specified in the current
setting.
354 This command is not available during calculation.
390 Command error.
391 Delimiter error.
392 Parameter error.
393 No permission.
394 No such connection.
395 Use “quit” to close this connection. Attempted to disconnect its own connection.
396 Failed to disconnect.
397 No TCP control block. The control block of the specified connection cannot be found.
400 Input username.
401 Input password.