Setup Operations for Convenient Functions (Setting Mode)
1
2
3
4
5
6
7
8
9
10
11
12
App
Index
6.13  Performing Calibration Correction (/CC1 
Option)
The values of the measuring input signal are corrected using specified segments, and 
the results are used as measured values.
If you change the input range, set the correction point (measured value) and correction 
values again.
Procedure
1.
  Hold down the 
MENU
 key for 3 seconds to enter Setting mode.
2.
  Press the   key or SHIFT +   key to select Calibration and then press the 
 key.
3.
  Set each item and press the   key.
Use the 
 key or SHIFT +   key to select values.
For the procedure on how to enter values or characters, see section 4.2.
If you press the ESC key in the middle of the operation, the settings entered up to 
then are cancelled, and the display returns to a higher level menu.
4.
  When the Setting complete screen appears, do either of the following:
Press the 
 key to set other channels.
If you are done, press the ESC key.
5.
  Hold down the 
MENU
 key for 3 seconds to return to Operation mode.
Set=Calibration
CH=01
01CH Calibration
Setting complete
Turn On or Off the calibration correction.
Set the first correction point 
(input value).
Select the channel.
(Displays the following when
 Calibration is On)
The new setting takes effect.
Yes: The setting is fixed.
No: The setting becomes invalid.
ESC/?
Calibration=On
1 datums= 0.000
Decision=Yes
Set the first correction value.
Likewise, set the second correction
point and correction value.
.
.
.
1 revise= 0.000