User Manual
IM 04P01B01-01E 
6-10
6.10  Setting the Alarm Delay Duration
This section explains the details of setting the duration of the delay alarm.
Procedure
1.
  Hold down the 
MENU
 key for 3 seconds to enter Setting mode.
2.
  Press the   key or SHIFT +   key to select Aux and then press the   key.
3.
  Press the   key or SHIFT +   key to select Alm delay T and then press the 
 key.
4.
  Set each item and press the   key.
Use the 
 key or SHIFT +   key to select values.
For the procedure on how to enter values or characters, see section 4.2.
If you press the ESC key in the middle of the operation, the settings entered up to 
then are cancelled, and the display returns to a higher level menu.
5.
  When the Setting complete screen appears, do either of the following:
Press the 
 key to set other channels.
If you are done, press the ESC key.
6.
  Hold down the 
MENU
 key for 3 seconds to return to Operation mode.
Set the channel range.
Set the alarm delay time.
The new setting takes effect.
ESC/?
First channel Last channel
Set=Aux
Aux=Alm delay T
CH=01-01
Duration= 10
01-01 Channel
Setting complete
Explanation
Duration
The duration can be set to a value in the range of 1 to 3600 s.
However, if the scan interval on the dot model is 2.5 s, the actual duration is an integer 
multiple of 2.5 s. If the set value is not equal to an integer multiple of 2.5 s, the duration is 
the smallest integer multiple of 2.5 s greater than the set value.
Example: The actual duration for 1 s and 2 s is 2.5 s.
<Related
 T
opics> Enabling the alarm delay function: Section 7.15