    Newest value output (GETNEW)
  Output the specified number of blocks (p4) 
of FIFO data back starting from the recent 
acquire position (block).
•  Parameters p2, p3, and p4 are valid when p1 
is set to GET or GETNEW.
•  If p4 is omitted, all the data of all blocks 
acquired in the FIFO buffer are output.
•  Be sure to set p2 and p3 so that p3 is greater 
than or equal to p2.
•  For the output flow of FIFO data, see appendix 
4.
IS  Outputs status information.
Syntax
  IS p1
<terminator>
p1
  Output status information (0)
Example
  Output status information.
IS0
Description The output status can be masked using the 
status filter (IF command).
FU  Outputs user information.
Syntax
  FU p1
<terminator>
p1
  Output user information (0)
Example
  Output user information.
FU0
Description Outputs the information of the user currently 
connected to the recorder
.
4.8  RS-422A/485 Dedicated 
Commands
ESC O  Opens the instrument.
The ASCII code of ESC is 1BH.
Syntax
  ESC O p1
<terminator:CR+LF>
p1
  Instrument address (01 to 32)
Example
  Open the instrument at address 01.
ESC O01
Description •
  Specifies the address of the instrument with 
which to communicate.
•
  Only one instrument can be opened at any 
given time.
•
  When an instrument is opened with the ESC 
O command, any other instrument that is 
currently open is automatically closed.
•  When this command is received correctly
, the 
recorder transmits the data “ESC 0 xx” (xx: 
Instrument’
s address). For the description of 
the response, see page 5-5.
•  Use CR+LF for the terminator. LF cannot be 
used.
ESC C  Closes the instrument.
The ASCII code of ESC is 1BH.
Syntax
  ESC C p1
<terminator:CR+LF>
p1
  Instrument address (01 to 32)
Example  Close the instrument at address 01.
ESC C01
Description •  Clears the current connection with the 
instrument.
•  When this command is received correctly, the 
recorder transmits the data “ESC C xx” (xx: 
Instrument’s address). For the description of 
the response, see page 5-5.
•  Use CR+LF for the terminator. LF cannot be 
used.