User Manual

Commands
1
2
3
4
5
6
App
Index
IM 04P01B01-17E
4-29
MC Clears the message printout
buffer.
Syntax
MC p1
<terminator>
p1
Clear the message printout buf
fer (0)
Example
Clear the message printout buf
fer.
MC0
VG Resets the report data of the
periodic printout.
Syntax
VG p1
<terminator>
p1
Operation type
2
Reset the report data of the periodic
printout.
Example
Reset the report data of the periodic printout.
VG2
Description
This setting is valid when the recorder is set to
print the report data using periodic printout.
YC Initializes settings.
Syntax
YC p1
<terminator>
p1
Initialization type
0
Initialize the Setting mode and Basic
Setting mode settings.
1
Initialize the Setting mode settings.
Example
Initialize the Setting mode and Basic Setting
mode settings.
YC0
UY Stops the record position
adjustment.
Syntax
UY p1
<terminator>
p1
Fixed to 0.
Query
UY?
Example
Stop the record position adjustment.
UY0
Description
Returns the execution status of the record
position adjustment in response to a query
.
0: Stopped, 1: In progress
4.7 Output Commands
BO Sets the byte output order.
Syntax
BO p1
<terminator>
p1
Byte order
0
Outputs the data MSB first.
1
Outputs the data LSB first.
Query
BO?
Example
Output the data MSB first.
BO0
Description
This command applies to the byte order of
numerical data during BINAR
Y
output.
CS Sets the checksum (/C3 option).
Syntax
CS p1
<terminator>
p1
Use/Not use the checksum
0
Not use
1
Use
Query
CS?
Example
Use the checksum.
CS1
Description
This command can be used only on the RS-
422A/485 communication interface.
IF Sets the status filter.
Syntax
IF p1
<terminator>
p1
Status filter value
(0.0.0.0 to 255.255.255.255)
Query
IF?
Example
Set the status filter value to 1.0.4.0.
IF1.0.4.0
Description
For details, see chapter 6.
Note
Initialization of BO/CS/IF Command Settings
For serial communications
Settings entered using the BO/CS/IF commands revert to their
initial values when the recorder is reset (when the recorder is
power cycled, or the user exits the basic setting mode with the
YE command).
Byte output order, checksum: 0
Status filter: 255.255.255.255
If the recorder is reset, you must restore these settings.
For
Ethernet communications
Settings entered using the BO/IF commands revert to their
initial values when the connection to the recorder is cut. After
reconnecting the recorder, you must reenter the settings.
4.6 Control Commands/4.7 Output Commands