User Manual
Commands
1
2
3
4
5
6
App
Index
IM 04P01B01-17E
4-27
  When p1 =
 FULL
    µR10000:  An integer between –30 to 30 
(reference value: 1000)
     µR20000:  An integer between –50 to 50 
(reference value: 1800)
Example  Adjust the 100% position. Set the record position 
adjustment value to 25.
UAFULL,25
Description •  C
arry out the adjustment in the following order: 
P1 = Hysteresis, ZERO, and then FULL.
•  Check the recorded result and correct the 
record position adjustment value.
•  To end the adjustment, execute the UY0 
command. If you do not end the adjustment, 
you cannot change the execution mode.
•
  The unit for p2 is dot (0.1 mm).
UG  Sets the Setting mode menu 
selection.
Syntax
  UG p1,p2,p3,p4,p5,p6,p7,p8,p9,p10 
<terminator>
p1
  Range (ON, OFF)
p2
  Bias (ON, OFF)
p3
  Alarm (ON, OFF)
p4
  Units (ON, OFF)
p5
  Chart speed (ON, OFF)
p6
  Other (ON, OFF)
p7
  Calibration correction (ON, OFF) (/CC1 
option) 
p8
  Computation (ON, OFF) (/M1 option) 
p9
  Batch name (ON, OFF) (/BT1 option) 
p10
 Batch details (ON, OFF) (/BT1 option) 
Query
  UG?
Example  Display all menus. 
UGON,ON,ON,ON,ON,ON,ON,ON,ON,ON
Description •  Parameter p7 can be specified on models with 
the calibration correction (/CC1 option).
•  Parameter p8 can be specified on models with 
the computation function (/M1 option). 
•
  Parameters p9 and p10 can be specified on 
models with the header printout (/BT1 option). 
Note
Parameters p7 and higher are skipped when no options are 
installed.
UH  Sets the FUNC key menu 
selection.
Syntax
  UH p1,p2,p3,p4,p5
<terminator>
p1
  Manual printout (ON, OFF)
p2
  Setup printout (ON, OFF)
p3
  Message printout (ON, OFF)
p4
  Buffer clear (ON, OFF)
p5
  Periodic printout (ON, OFF)
Query
  UH?
Example  Display all menus. 
UHON,ON,ON,ON,ON
UE  Selects enable/disable for 
Start printout/End printout and 
message format (/BT1 option).
When using Start printout/End printout 
Syntax
  UE p1,p2,p3,p4
<terminator>
p1
  Use of Start printout/End printout (USE)
p2
  Number of digits for lot number (4, 6, NOT)
p3
  Use/Not use Start printout 2/End printout 2 
(USE, NOT)
p4
  Use/Not use message format (USE, NOT)
Query 
UE?
Example  Perform Start printout/End printout, set 6 digits 
for lot number, use Start printout 2/End printout 2 
and message format. 
UEUSE,6,USE,USE
Description
 •  Set the lot number using the VH command. 
•  Set the message format using the VM 
command. 
When not using Start printout/End printout 
Syntax
  UE p1,p2
<terminator>
p1
  Start printout/End printout not used (NOT)
p2
  Use/Not use message format (USE, NOT)
Query
  UE?
Example  Use message format but do not use Start 
printout/End printout. 
UENOT,USE
Description Set the message format using the VM command. 
YE  Exits from Basic Setting mode.
When the YE command is executed, communication is 
dropped.
Syntax
  YE p1
<terminator>
p1
  Apply/Discard the settings   
STORE
  Apply the settings
ABORT
  Discard the settings
Example  Apply the basic settings.
YESTORE
Description Applies the settings changed with the basic 
setting commands.
XE  Exits from Basic Setting mode.
Communication is not dropped when the XE command is 
executed.
Syntax
  XE p1
<terminator>
p1
  Apply/Discard the settings
STORE
  Apply the settings
ABORT
  Discard the settings
Example  Apply the basic settings.
XESTORE
Description The settings specified by the following 
commands and saved using the XE command 
take effect after the recorder is power cycled.
  YS, YB, YA, YN, YD, YQ, and YK
4.5 Basic Setting Commands