User Manual
Commands
1
2
3
4
5
6
App
Index
IM 04P01B01-17E
4-25
p6
  Printout ON/OFF (ON, OFF)
Query
  XQ[p1]?
Example
  Set an absolute timer to timer number 1. Set the 
interval to 30 minutes, the reference time to hour 
7, reset the data when the timer expires, and 
disable printout.
XQ1,ABSOLUTE,30min,07,ON,OFF
Description
 Each time the interval specified by p3 elapses 
from the time specified by p4, the operations set 
with p5 and p6 are performed.
When using the relative timer
Syntax
  XQ p1,p2,p3,p4,p5
<terminator>
p1
  T
imer number (1 or 2)
p2
  Timer type (RELA
TIVE)
p3
  Interval (fixed to HH:MM format)
HH
  Hour (00 to 24)
MM
  Minute (00 to 59)
  Set the interval in the range 00:01 to 24:00
p4
  Reset/not reset the data when the timer 
expires (ON/OFF)
p5
  Printout ON/OFF (ON, OFF)
Query
  XQ[p1]?
Example
  Set a relative timer to timer number 1. Set the 
interval to 1 hour 15 minutes, reset the data 
when the timer expires, and disable printout.
XQ1,RELATIVE,01:15,ON,OFF
Description
 Each time the interval specified by p3 elapses 
from the time the computation is started, the 
operations set with p4 and p5 are performed.
UN  Changes the assignment of 
channels to the recording pen 
(pen model, /M1 option).
Syntax
  UN p1,p2
<terminator>
p1
  Pen number (1 to 4)
p2
  Channel number
Query
  UN[p1]? 
Example
  Assign computation channel 0A to pen number 4.
UN4,0A
US  Sets the computation error 
procedure (/M1 option).
Syntax
  US p1,p2
<terminator>
p1
  Procedure taken when an error occurs.
+OVER
  Handle error data as +OVER
−OVER
  Handle error data as –OVER
p2
  Procedure taken when the “input over” is 
detected (procedure taken when an “input 
over” value is applied to TLOG.SUM or 
TLOG.AVE computation)
SKIP
  Do not use the data for 
computation.
LIMIT
  Use the limit value given by the 
recorder for computation. 
Query
  US? 
Example  When the computed result is in error, handle the 
error data as +OVER. Use the limit value given 
by the recorder for computation instead of the 
“input over” value. 
US+OVER,LIMIT
YB  Sets the host name and domain 
name (/C7 option).
Syntax
  YB p1,p2
<terminator>
p1
  Host name (up to 64 characters)
p2
  Domain name (up to 64 characters)
Query
  YB?
Example
  Set the host name to “ABC” and the domain 
name to “recorder.co.jp.”
YBABC,recorder.co.jp
Description
 The settings specified by this command and 
saved using the XE command take ef
fect after 
the recorder is power cycled.
YA  Sets the IP address (/C7 option).
Syntax
  YA p1,p2,p3
<terminator>
p1
  IP
 address (0.0.0.0 to 255.255.255.255)
p2
  Subnet mask
  (0.0.0.0 to 255.255.255.255)
p3
  Default gateway
  (0.0.0.0 to 255.255.255.255)
Query
  YA?
Example
  Set the IP
 address to 192.168.111.24, the subnet 
mask to 255.255.255.0, and the default gateway 
to 0.0.0.0.
YA192.168.111.24,255.255.255.0, 
0.0.0.0
Description
 The settings specified by this command and 
saved using the XE command take ef
fect after 
the recorder is power cycled.
YN  Sets the DNS (/C7 option).
When not using the DNS
Syntax
  YN p1
<terminator>
p1
  Use/Not use the DNS (OFF)
Query
  YN?
Example  Do not use the DNS.
YNOFF
When using the DNS
Syntax
  YN p1,p2,p3,p4,p5
<terminator>
p1
  Use/Not use the DNS (ON)
p2
  Address of the primary DNS server
  (0.0.0.0 to 255.255.255.255)
p3
  Address of the secondary DNS server
  (0.0.0.0 to 255.255.255.255)
p4
  Domain suffix 1 (up to 64 characters)
p5
  Domain suffix 2 (up to 64 characters)
Query
  YN?
Example  Use the DNS server at 192.168.0.1.
YNON,192.168.0.1
4.5 Basic Setting Commands