User Manual
Commands
1
2
3
4
5
6
App
Index
IM 04P01B01-17E
4-15
Example Set the display (VFD) brightness to 2 and the
internal illumination to 1.
VF2,1
Description
The brightness increases as the value increases.
TD Sets the DST.
Syntax
TD p1,p2,p3,p4,p5,p6,p7,p8,p9
<terminator>
p1
Use/Not use DST (USE, NOT)
p2
DST start time: month (Jan, Feb, Mar, Apr,
May, Jun, Jul, Aug, Sep, Oct, Nov, or Dec)
p3
DST
start time: number of the week in the
month (1st, 2nd, 3rd, 4th, or Last)
p4
DST
start time: day of the week (Sun, Mon,
Tue, Wed, Thu, Fri, or Sat)
p5
DST
start time: hour (0 to 23)
p6
DST end time: month (Jan, Feb, Mar, Apr,
May, Jun, Jul, Aug, Sep, Oct, Nov, or Dec)
p7
DST
end time: number of the week in the
month (1st, 2nd, 3rd, 4th, or Last)
p8
DST
end time: day of the week (Sun, Mon,
Tue, Wed, Thu, Fri, or Sat)
p9
DST
end time: hour (0 to 23)
Query
TD[p1]?
Example
Enable DST
and set the DST
start time to hour
0 on the 2nd Sunday of June and the DST end
time to hour 0 on the 2nd Sunday of December.
TDUse,Jun,2nd,Sun,0,Dec,2nd,Sun,0
Description
•
The DST
start time and end time cannot be
set to the same time.
•
The
TD command can be used on the
µR10000 and µR20000 with system version
1.1
1 or later. On the µR10000 with system
version 1.02 or earlier, use the SS command.
SS Sets the DST.
Syntax
SS p1,p2,p3
<terminator>
p1
Standard time/DST
selection
Winter
Standard time
Summer
DST
p2
Switch time setting ON/OFF (ON, OFF)
p3
Date and time (fixed to the YY/MM/DD HH
format)
YY
Y
ear (00 to 99, the lower 2 digits)
MM
Month (01 to 12)
DD
Day (01 to 31)
HH
Hour (00 to 23)
Query
SS[p1]?
Example
Switch to DST
on 0 hour of June 1st, 2005.
SSSummer,ON,05/06/01 00
Description
•
Parameter p3 is not available when p2 is OFF
.
• The SS command can be used on the
µR10000 with system version 1.02 or earlier.
SO Sets the computing equation
(/M1 option).
Syntax
SO p1,p2,p3,p4,p5,p6
<terminator>
p1
Computation channel number
p2
Turn ON/OFF the computing equation (ON,
OFF)
p3
Computing equation (up to 120 characters)
p4
Leftmost value of span (−9999999 to
99999999)
p5
Rightmost value of span (−9999999 to
99999999)
p6
Span decimal point position (0 to 4)
Query
SO[p1]?
Example Calculate the sum of channel 01 and 02 on
computation channel 0A. Set the leftmost and
rightmost values of span to –10.00 and 15.00,
respectively.
SO0A,ON,01+02,−1000,1500,2
Description • This command cannot be specified while
computation is in progress.
• For a description on the computing equations,
see the µR10000 Recorder User’s Manual or
the µR20000 Recorder User’s Manual.
• For p4 and p5, enter a value using 7 digits
or less ,excluding the decimal, for negative
numbers and 8 digits or less for positive
numbers.
SK Sets the computation constant
(/M1 option).
Syntax
SK p1,p2
<terminator>
p1
Constant number (K01 to K30)
p2
Constant (up to 11 characters)
The range is –9.9999E+29 to –1.0000E–30,
0, 1.0000E–30 to 9.9999E+29.
(The + sign of “E+” can be omitted.)
Query
SK[p1]?
Example Set 1.0000E−10 to constant K01.
SKK01,1.0000E−10
Description This command cannot be specified while
computation is in progress.
SJ Sets the timer used in TLOG
computation (/M1 option).
Syntax
SJ p1,p2,p3
<terminator>
p1
Computation channel number
p2
Timer number (Periodic, 1, 2)
p3
Sum scale (for TLOG.SUM)
OFF
Integrates the data per scan interval.
/s
Integrates the physical amount that
are in unit of /s.
/min
Integrates the physical amount that
are in unit of /min.
/h
Integrates the physical amount that
are in unit of /hour.
/day
Integrates the physical amount
that are in unit of /day.
4.4 Setting Commands