User Manual
Commands
1
2
3
4
5
6
App
Index
IM 04P01B01-17E
4-13
SC  Sets the chart speed.
Syntax
  SC p1
<terminator>
p1 
Chart speed
Query
  SC?
Example  Set the chart speed to 25 mm/h.
SC25
Description Select the chart speed from the list of choices 
below.
Pen model
5 to 12000 mm/h (82 levels, see section 4.3)
Dot model
1 to 1500 mm/h (1 mm steps)
SD  Sets the date and time.
Syntax
  SD p1
<terminator>
p1
  Date and time (fixed to the 
YY/MM/DD_HH:
MM:SS format)
YY
  Y
ear (00 to 99, the lower 2 
digits)
MM
  Month (01 to 12)
DD
  Day (01 to 31)
_
  Space
HH
  Hour (00 to 23)
MM
  Minute (00 to 59)
SS
  Second (00 to 59)
Query
  SD?
Example
  Set the recorder clock to 13:00:00, December 1, 
2004.
SD04/12/01 13:00:00
Description
 •
  The format of p1 is fixed to 17 characters. 
An 
error results if a space is inserted in positions 
other than the specified positions.
•
  On the µR10000 with system version 1.02 or 
earlier
, the parameter p2 can be used.
p2
 Standard time or DST
 (winter
, summer)
Winter
  Standard time
Summer
  DST
VT  Sets the dot printing interval 
(dot model).
Syntax
  VT p1
<terminator>
p1 
Dot printing interval setting.
AUTO
  Automatically adjust the dot 
printing interval according to 
the chart speed.
FIX
  Record at the fastest printing 
interval.
Query
  VT?
Example
  Record at the fastest printing interval.
VTFIX
Description
 When trend recording is performed on 6 
channels, the fastest printing interval is 10 
seconds.
SZ  Sets zone recording.
Syntax
  SZ p1,p2,p3
<terminator>
p1 
Channel number
p2 
Leftmost position of the zone (µR10000: 0 to 
95, µR20000: 0 to 175) [mm]
p3 
Rightmost position of the zone (µR10000: 5 
to 100, µR20000: 0 to 180) [mm]
Query
  SZ[p1]?
Example  Display channel 02 in a zone between 30 and 50 
mm.
SZ02,30,50
Description •  Be sure that p3 is greater than p2 and that the 
zone width (p3 – p2) is greater than or equal 
to 5 mm.
•  On models with the computation function 
(/M1 option), computation channels can be 
specified.
SP  Sets the partial expanded 
recording.
Syntax
  SP p1,p2,p3,p4
<terminator>
p1 
Channel number
p2 
Partial expanded recording ON/OFF (ON, 
OFF)
p3 
Boundary position (1 to 99) [%]
p4 
Boundary value
Query
  SP[p1]?
Example  Set the 25% position of channel 01 to 1.000 V.
SP01,ON,25,1000
Description
 •
  When the input range is set to SKIP
 (SR 
command), p2 cannot be turned ON.
•
  Set p3 as a percentage where 100 mm is 
assumed to be 100%.
•  P
a
rameter p4 can be set in the range (leftmost 
value of span + 1) to (rightmost value of span 
– 1). If scaling is enabled, the range is (leftmost 
value of scaling – 1) to (rightmost value of 
scaling + 1). Set the parameter using an 
integer.
•  This setting is possible when partial expanded 
recording is enabled (UK command).
•  This command cannot be used if the partial 
expanded recording range does not exist 
(when the span width is set to 1, for example).
•  On models with the computation function 
(/M1 option), computation channels can be 
specified.
•  When the computation equation setting (SO 
command) is turned OFF, p2 cannot be turned 
ON.
4.4 Setting Commands