User Manual
4-10
IM 04P01B01-17E
4.4 Setting Commands
4.4  Setting Commands
SR  Sets the input range.
When setting channels to skip
Syntax
  SR p1,p2
<terminator>
p1
  Channel number
p2
  Measurement mode (SKIP)
Query
  SR[p1]?
Example
  Set channel 01 to skip.
SR01,SKIP
Description
 •
  This command cannot be specified while 
computation is in progress.
•
  Measurements are not made on channels that 
are set to SKIP
.
When setting channels to voltage, TC, RTD, or 
ON/OFF input
Syntax
  SR p1,p2,p3,p4,p5
<terminator>
p1
  Channel number
p2
  Measurement mode (Input type)
VOLT
  DC voltage
TC
  Thermocouple
RTD
  Resistance temperature 
detector
DI
  ON/OFF input
p3
  Range type
p4
  Leftmost value of span
p5
  Rightmost value of span
Query
  SR[p1]?
Example  Measure 0°C to 1760.0°C on channel 01 using 
thermocouple type R.
SR01,TC,R,0,17600
Description •  This command cannot be specified while 
computation is in progress.
•  Set p3 according to the table in section 4.3.
•  For p4 and p5, enter an integer value of 5 
digits or less according to the table in section 
4.3. The decimal position is fixed to the 
position indicated in the table in section 4.3.
When setting channels to 1-5V range
Syntax
  SR p1,p2,p3,p4,p5,p6,p7,p8
<terminator>
p1
  Channel number
p2
  Measurement mode (Input type) (1-5V)
p3
  Leftmost value of span
p4
  Rightmost value of span
p5
  Leftmost value of scaling (−20000 to 30000)
p6
  Rightmost value of scaling (−20000 to 
30000)
p7
  Scale decimal point position (0 to 4)
p8
  Whether to use 1-5V low-cut function (ON, 
OFF)
Query
  SR[p1]?
Example  Set channel 01 to 1-5V range and scale the input 
value in the range 0.0 to 1200.0. Do not use the 
low-cut function.
SR01,1-5V,1000,5000,0,12000,1,OFF
Description •  This command cannot be specified while 
computation is in progress.
•  For p3 and p4, enter an integer value of 5 
digits or less according to the table in section 
4.3. The decimal position is fixed to the 
position indicated in the table in section 4.3.
•
  Be sure that p6 is greater than p5.
•  Parameter p8 is valid only when the low-cut 
function is enabled (see the UF command).
When computing the difference between channels
Syntax
  SR p1,p2,p3,p4,p5
<terminator>
p1
  Channel number
p2
  Measurement mode (DELTA)
p3
  Reference channel
p4
  Leftmost value of span
p5
  Rightmost value of span
Query
  SR[p1]?
Example  Set channel 03 to channel difference 
computation with respect to channel 01 
(reference channel). Set the leftmost and 
rightmost values of span to −200.0 and 200.0, 
respectively.
SR03,DELTA,01,−2000,2000
Description •  This command cannot be specified while 
computation is in progress.
•  The reference channel must be a channel that 
is smaller in channel number than itself.
•
  The range type is the same as that of the 
reference channel.
•  For p4 and p5, enter an integer value of 5 
digits or less according to the table in section 
4.3. The decimal position is fixed to the 
position indicated in the table in section 4.3.
When setting the linear scaling 
Syntax
  SR p1,p2,p3,p4,p5,p6,p7,p8,p9 
<terminator>
p1 
Channel number
p2 
Measurement mode (SCALE)
p3 
Input type
VOLT
  DC voltage
TC
  Thermocouple
RTD
  Resistance temperature 
detector
DI
  ON/OFF input
p4
  Range type
p5
  Leftmost value of span
p6
  Rightmost value of span
p7
  Leftmost value of scaling (−20000 to 30000)
p8
  Rightmost value of scaling (−20000 to 
30000)
p9
  Scaling decimal point position (0 to 4)
Query
  SR[p1]?
Example  Scale channel 02 whose input range is 0 to 10 V 
to −100.0 to 500.0.
SR02,SCALE,VOLT,20V,0,1000,−1000, 
5000,1