User Manual
47
IM 04P01B01-02E 
FUNC Key Operations in Operation Mode
The operations below can be carried out with the FUNC key in Operation mode. 
References to the µR10000 Recorder User’s Manual (IM 04P01B01-01E) provided on 
the CD-ROM are given in parentheses.
:
Use the  key.
key.
Message 1
Message 2
Message 3
Message 4
Message 5
Alarm
Password
Message
Start/Stop
Reset
Reset
R. exchange = End
List Start/List Stop
Setup Start/Setup Stop
ManualStart/Manual Stop
:
Use the 
Alarm ACK
Message
Buffer clear
Keylock
Periodic
Print out
R. exchange
(section 3.12)
(section 9.1)
(section 3.10)
(sections 3.7, 3.8)
(sections 3.9, 3.10)
(section 3.13)
(section 3.11)
(section 3.4)
Pen exchange = End
Pen exchange
(section 3.3)
Math
key
FUNC
Alarm output release operation. This is displayed when the 
settings are changed to use the alarm ACK operation. 
Starts/stops the computation. This is displayed on models with 
the computation function (/M1 option). 
Resets the computation. 
Executes manual printout (printout of measured values). 
Executes the setting (Setting mode) printout.
Executes setting (Basic Setting mode) printout.
Prints message 1.
Prints message 2.
Prints message 3.
Prints message 4.
Prints message 5.
Clears the data waiting to be printed in the alarm printout buffer.
Clears the data waiting to be printed in the message printout buffer.
This is displayed when the settings are changed to use the key 
lock function. It is used to activate or release the key lock.
Resets the computed value when printing of the report data 
(average, etc.) is specified in periodic printout and restarts the 
calculation of the report data from that point.
Moves the recording pen to a position that is easily accessible for 
replacement on the pen model.
Moves the printer carriage near the center position when replacing 
the ribbon cassette on the dot model.
Recorder version: 1.31
Printing Measured Values (Manual Printout)
The measured values of all channels are printed.
Starting the Manual Printout
1.
  Press the 
FUNC
 key.
The Func= screen appears.
2.
  Press the   key to select Print out and then press the   key.
Func=Print out
3.
  Press the   key with ManualStart shown on the screen.
Manual printout starts. The screen returns to the data display screen.
Print=ManualStart
Note
•  When manual printout is executed, trend recording is suspended. However, the recorder 
continues the measurement and alarm detection (in the background).
•  When manual printout is complete, trend recording resumes.
•  If an alarm occurs during the manual printout, the alarm is printed after the recording 
resumes.
Recording/Displaying Data
*  When the key or key is pressed while holding down the SHIFT key, the operation is reversed as 