User Manual
20
IM 04P01B01-02E 
Wiring Procedure (Panel Mount Type)
The power supply terminals and protective ground terminals are located on the rear 
panel.
1.
  Turn OFF the power switch on the recorder and open the power terminal cover.
2.
  Wire the power cord and the protective ground cord to the power supply terminals.
Use ring-tongue crimp-on lugs (designed for 4-mm screws).
3.
  Close the power supply terminal cover and secure it with the screw.
The proper torque for tightening the screws is 0.6 N•m.
Screw for fixing 
the power terminal 
cover in place
Power
terminal block
Power
terminal cover
Open
Philips
screwdriver
Crimp-on lug 
with insulation 
sleeves
Power cord
Protective ground 
cord
• Other than /P1 • /P1
 (24 VDC/AC power supply)
L
N
L
N
DC/AC24V
+
-
Wiring Procedure (Portable Type (/H5x Option))
1.
  Check that the power switch to the instrument is turned OFF.
2.
  Connect the plug on the accessory power cord to the power supply connector on 
the rear panel.
100-240V~
 50/60Hz
40VA MAX
3.
  Ensure that the power outlet to be used meets the conditions on the previous 
page and that the voltage of the power supply is within the maximum voltage 
rating of the power cord, then connect the other end of the power cord to the 
power supply outlet.
The AC outlet must be of a three-prong type with a protective earth ground termi-
nal.
Turning ON/OFF the Power Switch
The power switch is located inside the door at the lower right. The power switch is a push 
button.
Press once to turn it ON and press again to turn it OFF. When the power switch is turned 
ON, a self-diagnosis program runs for a few seconds, and the recorder is ready for 
operation.
Installing/Wiring the Recorder