User Manual
4-3SM 04P02B01-01E
Adjustments
4
● Procedure
The following explains the procedure when working with channel 1.
1. Press the 
MENU
 and 
CH UP
 keys while turning ON the power. Menu = A/D
Adjust is displayed.
2. With Menu = A/D Adjust is displayed, press the 
 key.
Menu=A/D Adjust
3. Input the password (see page 3-5).
Password=
4. Select Mode = Cal/Exec, then press the   key.
Mode=Cal/Exec
5. Select A/D No. = 1, then press the   key.
* For channel 3, select A/D No. = 2.
A/D No.=1
6. Select Range = 20 mVZ.
Range=20mVZ
7. Enter the zero value for the 20 mV range, then press the   key.
After measurement, the adjusted value is displayed.
Range=20mVZ
Zero:-200
8. Press the   key.
You are returned to the A/D No. = screen.
9. Leave the A/D No. unchanged, then press the   key.
A/D No.=1
10. Select Range = 20 mVF, then press the   key.
Range=20mVF
11. Enter the Full value for the 20 mV range on channel 1, then press the   key.
After measurement, the adjusted value is displayed.
Range=20mVF
Full:32768
12. Repeat steps 5 to 11 to adjust the following ranges: 60 mV, 200 mV, 1 V, 2 V, 6
V, 20 V, Pt100, Cu10 (with the /N1 option).
Saving Adjusted Values
13. When all adjustments are finished, select Mode = End, then press the 
 key.
Mode=End
14. Select A/D No. = 1, then press the   key.
* You can select an A/D No. of 1, 2, or 1+2. The adjusted value is saved on the AD
converter of the selected number.
A/D No.=1
4.2 Adjustments to the Pen Model
* Select using the key. To enter numerical values, move the cursor to the entry point using the 
   key. Select the value using the key. You can hold down the SHIFT key while pressing 
   key or key to reverse the key’s action.