User Manual
3-13SM 04P02B01-01E
Testing
3
Keys
● Procedures
1. Turn ON the power switch while pressing the 
MENU
 and 
CH UP
 keys. Menu =
A/D Adjust is displayed.
2. Select Menu = Test, and press the 
 key.
3. Choose Test = Key and press the   key.
Confirm that the appropriate box turns On or OFF when each key is pressed.
The box of the display corresponds to the order of the keys.
Press the 
ESC/?
 key twice to return to the Test = Key screen.
Enter 2 ESC, Out.
* Select using the key. You can hold down the SHIFT key while pressing key to reverse the 
  key’s action.
Power Supply Frequency Detection
● Procedures
1. Turn ON the power switch while pressing the 
MENU
 and 
CH UP
 keys. Menu =
A/D Adjust is displayed.
2. Select Menu = Test, and press the 
 key.
3. Choose Test = Frequency and press the   key.
The detected power supply frequency is displayed. Confirm that it matches the
actual power supply frequency.
Frequency
50.0Hz
* Select using the key. You can hold down the SHIFT key while pressing key to reverse the 
  key’s action.
Battery Backup
● Procedures
1. Set the date and time on the recorder. Also, set an arbitrary channel to the 20
mV range.
2. Turn OFF the power switch.
3. Turn ON the power after approximately one minute, then check that the set
date, time, and range are not initialized.
3.4 Test Procedures