User Manual

3-10 SM 04P02B01-01E
Procedures
Warm up the instrument for thirty minutes before performing the test. Set the ranges
on the recorder, and input the values determined for each range. Check whether the
displayed measured values meet the criteria.
DC voltage generator
To the input
terminals
+
Variable resistors
The resistance of three
lead wires must be equal.
L
N
Criteria
Range Input Range Allowed by Test Accuracy Specification Remarks
20mV +20 mV ±3 digits ±(0.1% of rdgs + 2digits)
0 mV ±1 digits
20 mV ±3 digits
60mV +60 mV ±6 digits
0 mV ±1 digits
60 mV ±6 digits
200mV +200 mV ±3 digits
0 mV ±1 digits
200 mV ±3 digits
2V +2 V ±3 digits
+1 V ±2 digits
0 V ±1 digits
1 V ±2 digits
2 ±3 digits
6V +6 V ±6 digits
0 V ±1 digits
6 V ±6 digits
20V +20 V ±3 digits
0 V ±1 digits
20 V ±3 digits
50V +50 V ±5 digits ±(0.1% of rdgs + 3 digits)
0 V ±2 digits
50 V ±5 digits
Pt100 200°C: 18.52 ±0.4°C ±(0.15% of rdgs + 0.3°C)
0°C: 100 ±0.2°C
600°C: 313.71 ±0.9°C
TC-T 0 mV ±0.2°C EXT RJC mode
TC-K 0 mV ±0.3°C EXT RJC mode
Cu10 200°C: 1.326 ±1.4°C ±(0.4% of rdgs + 1.0°C) Performed when the /N1 option
(GE) 0°C: 9.036 ±0.8°C is specified.
300°C: 20.601 ±1.7°C
Pt50 200°C: 8.57 ±0.8°C ±(0.3% of rdgs + 0.6°C) Performed when the /N3 option
0°C: 50 ±0.4°C is specified.
600°C: 158.64 ±0.9°C
3.4 Test Procedures