User Manual
4-8 SM 04P01B01-01E
● Procedure
1. Press the 
MENU
 and 
CH UP
 keys while turning ON the power. Menu = A/D Adjust
is displayed.
2. With Menu = A/D Adjust is displayed, press the 
 key.
Menu=A/D Adjust
3. Input the password (see page 3-6).
Password=
4. Select Mode = Cal/Exec, then press the   key.
Mode=Cal/Exec
5. Select A/D No. = 1, then press the   key.
A/D No.=1
6. Select Range = 20 mV.
Range=20mV
7. Enter the zero value for the 20 mV range on channel 1 and the Full value on
channel 2, then press the   key.
After measurement, the adjusted value is displayed.
Range=20mV
 Z:-585 F:32769
8. Press the   key.
You are returned to the A/D No. = screen.
9. Repeat steps 5 to 8 to adjust the following ranges: 60 mV, 200 mV, 1 V, 2 V, 6
V, 20 V, Pt100, Cu10 (with the /N1 option).
Saving Adjusted Values
10. When all adjustments are finished, select Mode = End, then press the 
 key.
Mode=End
11. Select A/D No. = 1, then press the   key.
A/D No.=1
12. With End = Store displayed, press the   key.
The adjusted value is saved to the A/D converter.
End=Store
Concluding Adjustments
13. Select Menu = End, then press the   key. You are returned to the operation
mode.
Menu=End
4.3 Adjustments to the Dot Model
* Select using the key. To enter numerical values, move the cursor to the entry point using the 
   key. Select the value using the key. You can hold down the SHIFT key while pressing 
   key or key to reverse the key’s action. 