User Manual

SM 04P01B01-01E 4-7
Adjustments
4
4.3 Adjustments to the Dot Model
Adjusting the A/D Converters
Overview
Enter the value determined for each range, perform measurement, then write the
adjusted values to the A/D converter.
Instruments Used
Instrument Specifications
DC voltage generator Accuracy: ±0.005% of setting + 1 µV or better
Resistors 10 , 50 , 100 , 150 , 300 , Accuracy: 0.01% or better
Channel
Enter the Zero value for channel 1 and the Full value for channel 2, and perform the
adjustment.
Procedure
Connection
UUA
When DC range
DC Voltage
standard
DECADE
RESISTANCE
BOX
Channel 1
H
L
UUA
When RTD or Cu range
Channel 1
b
DC Voltage
standard
DECADE
RESISTANCE
BOX
Channel 2
H
L
Channel 2
+
+
B
A
b
B
A
(Zero value) (Zero value)
(Full value) (Full value)
UUA: Unit Under Adjustment
Input Value
Same as the pen model. See page 4-2.