User Manual
3-16 SM 04P01B01-01E
Ink Bleed (Dot Model)
● Procedures
1. Turn ON the power switch while pressing the 
MENU
 and 
CH UP
 keys.  Menu = A/D
Adjust is displayed.
2. Select Menu = Test, and press the 
 key.
3. Choose Test = Print and press the   key.
4. Select Menu = Ribbon, and press the 
 key.
Printing of six color test patterns begins. Confirm that the printed colors did not
bleed.
Purple Black Green  Blue  Brown Red
* Select using the key. You can hold down the SHIFT key while pressing key to reverse the 
  key’s action. 
Overscale Amount
● Overview
Pen Model
Check the accuracy of all pens. Use standard pens (or equivalent).
*A “standard pen” is one whose tip does not fluctuate more than ±0.1 mm.
Dot Model
Check an arbitrary channel.
● Recorder Settings
Set the input to the 2 V range, and −2.000 V to 2.000 V.
● Procedures
Input each value and record. Confirm that the recording position meets the criterion
below.
Criteria
Input Value Criteria
−2.3 V That it was 0.3 mm or more on the left side relative to the Zero recording position
2.3 V That it was 0.3 mm or more on the right side relative to the Full recording position
Relay Contact Output
● Overview
This test performed when the /A1, /A2, /A3, or /F1 option is specified. It is performed
for all relay contact outputs.
● Procedures
1. Turn ON the power switch while pressing the 
MENU
 and 
CH UP
 keys.  Menu = A/D
Adjust is displayed.
2. Select Menu = Test, and press the 
 key.
3. Choose Test = Alarm and press the   key.
The relay number selection screen (Relay=IO1) appears.
4. Select a relay number and check the relay action according to the table below.
* Select using the key. You can hold down the SHIFT key while pressing key to reverse the 
  key’s action. 
3.4 Test Procedures