User Manual

3-16 SM 04P01B01-01E
Ink Bleed (Dot Model)
Procedures
1. Turn ON the power switch while pressing the
MENU
and
CH UP
keys. Menu = A/D
Adjust is displayed.
2. Select Menu = Test, and press the
key.
3. Choose Test = Print and press the key.
4. Select Menu = Ribbon, and press the
key.
Printing of six color test patterns begins. Confirm that the printed colors did not
bleed.
Purple Black Green Blue Brown Red
* Select using the key. You can hold down the SHIFT key while pressing key to reverse the
key’s action.
Overscale Amount
Overview
Pen Model
Check the accuracy of all pens. Use standard pens (or equivalent).
*A “standard pen” is one whose tip does not fluctuate more than ±0.1 mm.
Dot Model
Check an arbitrary channel.
Recorder Settings
Set the input to the 2 V range, and 2.000 V to 2.000 V.
Procedures
Input each value and record. Confirm that the recording position meets the criterion
below.
Criteria
Input Value Criteria
2.3 V That it was 0.3 mm or more on the left side relative to the Zero recording position
2.3 V That it was 0.3 mm or more on the right side relative to the Full recording position
Relay Contact Output
Overview
This test performed when the /A1, /A2, /A3, or /F1 option is specified. It is performed
for all relay contact outputs.
Procedures
1. Turn ON the power switch while pressing the
MENU
and
CH UP
keys. Menu = A/D
Adjust is displayed.
2. Select Menu = Test, and press the
key.
3. Choose Test = Alarm and press the key.
The relay number selection screen (Relay=IO1) appears.
4. Select a relay number and check the relay action according to the table below.
* Select using the key. You can hold down the SHIFT key while pressing key to reverse the
key’s action.
3.4 Test Procedures