User Manual

3-15SM 04P01B01-01E
Testing
3
Dead Band (Pen Model)
Procedures
Check the accuracy of all pens. Use standard pens (or equivalent).
*A “standard pen” is one whose tip does not fluctuate more than ±0.1 mm.
1. Turn ON the power switch while pressing the
MENU
and
CH UP
keys. Menu = A/D
Adjust is displayed.
2. Select Menu = Test, and press the
key.
3. Choose Test = Dead Band and press the
key.
The dead band measurement pattern is automatically recorded near 0%, 50%,
and 100%. Measure the following portions of the recorded results using a Glass
scale (or equivalent), then confirm that the results meet the criteria.
Criterion: 0.2 mm or less
d
* Select using the key. You can hold down the SHIFT key while pressing key to reverse the
key’s action.
Pen Offset (Pen Model)
Procedures
Check all pens. Use standard pens (or equivalent).
*A “standard pen” is one whose tip does not fluctuate more than ±0.1 mm.
1. Turn ON the power switch while pressing the
MENU
and
CH UP
keys. Menu = A/D
Adjust is displayed.
2. Select Menu = Test, and press the
key.
3. Choose Test = Dead Band and press the
key.
The dead band measurement pattern is automatically recorded near 0%, 50%,
and 100%.
Measure the following portions of the recorded results using a Glass scale (or
equivalent), then confirm that the results meet the criteria.
Criteria
Item Criteria
A 2.2 ±1 mm
B 4.4 ±1 mm
C 6.6 ±1 mm
A B C
* Select using the key. You can hold down the SHIFT key while pressing key to reverse the
key’s action.
3.4 Test Procedures