User Manual

4-6
IM MX100-01E
Display
LED: Ethernet communication status (LINK, ACT)
Two-digit 7-segment LED: Status of the MX100 (unit number, operation status, and error
indication)
Communications
Interface: Ethernet (10BASE-T/100BASE-TX)
Connector type: RJ-45
Protocol: TCP, IP, UDP, ARP, and ICMP
Transmission function: Transmits measured values and settings
Reception function: Receives settings
TCP connection timeout: If there is no packet exchange for three minutes or more (not
congurable) when using TCP, a connection timeout occurs.
This mainly occurs when using the MX100/DARWIN API (sold
separately).
Other Functions
Time function: Normally operates using the PC clock
Backup function available (records the times of the data using
time stamps of the main module when communication is
disconnected)
Internal clock accuracy: ±100 ppm
Dip switches:
Used to initialize settings, etc. (number of switches: 8)
Access stop switch (CF card control switch):
Closes the le to which data is being written on the CF card
Saves the log le
Memory backup Settings and the clock are maintained by the internal lithium
battery, which has a lifetime of approximately 10 years (under
standard operating conditions)
General Specifications
Power consumption: Approximately 8 W
Insulation resistance: 20 MΩ or more at 500 VDC between the power supply terminal
and earth terminal
Withstand voltage: 1500 VAC (50/60 Hz) for one minute between the power supply
terminal and earth terminal
External dimensions: Approximately 92 (W) × 131 (H) × 137 (D) mm
Weight: Approximately 0.85 kg
External Dimensions
92 136.7
131
Unit: mm
If not specified, the tolerance is ±3%. However, in cases of less than 10 mm, the tolerance is ±0.3
mm.
4.2 Main Module (MX100-E) Specifications