User Manual

4-2
IM MX100-01E
4.2 Main Module (MX100-E) Specifications
Measurement
Style number: S3
Meas. range and accuracy:
See the measurement range and accuracy given in the
specications of each input module.
Max. number of inputs: 60 (however a maximum of six modules can be controlled)
Meas. interval: Select from 10, 50, 100, 200, 500 ms, 1, 2, 5, 10, 20, 30, or 60 s.
Up to three intervals dened for the input modules can be set
(multi-interval)
Synch. between modules:
Synchronized within the same sampling interval (in the same unit)
Synch. between channels:
4-CH, High-Speed Universal Module and Digital Input Module:
sync
hronized between channels. 10-CH, Medium-Speed Universal
Input Module, 30-CH, Medium-Speed DCV/TC/DI Input Module,
6-CH, Medium-Speed Four-Wire RTD Resistance Input Module, and
the 4-CH, Medium-Speed Strain Input Module: due to the scanner
type, not synchronized between channels
T
ime setting: Congured based on the time information sent from the PC at the
start of measurement
Filter function: First
-order lag lter, applicable measurement types:
DC voltage, thermocouple, RTD, strain, and resistance. You can select
a time constant (time until 63.2% of the output value is reached)
corresponding to the sampling interval indicated in the table below.
Time constant = measurement interval × N (where N = 5, 10, 20,
25, 40, 50, or 100)
Measurement
Interval (s)
Selectable Time Constants (s)
N = 5 N = 10 N = 20 N = 25 N = 40 N = 50
N = 100
0.01 0.05 0.1 0.2 0.25 0.4 0.5 1
0.05 0.25 0.5 1 1.25 2 2.5 5
0.1 0.5 1 2 2.5 4 5 10
0.2 1 2 4 5 8 10 20
0.5 2.5 5 10 12.5 20 25 50
1 5 10 20 25 40 50 100
2 10 20 40 50 80 100 200
5 25 50 100 125 200 250 500
10 50 100 200 250 400 500 1000
20 100 200 400 500 800 1000 2000
30 150 300 600 750 1200 1500 3000
60 300 600 1200 1500 2400 3000 6000
Computation
Computation performed on the main module
Difference comp. between channels:
Difference computation between arbitrary channels (DC voltage,
thermocouple, RTD, DI, strain, resistance, and scaling)
Linear scaling: Possi
ble scaling ranges: DC voltage, thermocouple, RTD, DI, strain,
and resistance
Scaling range: –30000 to 30000
Scaling display range: –32000 to 32000
Decimal point position: Arbitrary