User Manual
4-1
IM MX100-01E
Specification
3
2
1
4
5
Index
Chapter 4 Specification
4.1 Common Specifications
Standard operating conditions
Operating temperature range: 0 to 50°C
Operating humidity range: 20 to 80% RH for 0 to 40°C
10 to 50% RH for 40 to 50°C
Rated supply voltage: AC power supply, 100 to 240 VAC
Supply voltage range used: AC power supply, 90 to 250 VAC
Power supply frequency: 50 Hz ±2%, 60 Hz ±2%
Power consumption: Up to approximately 70 VA maximum when six modules
are used
Vibration: 10 to 60 Hz 0.2m/s
2
or less
Shock: Not allowed
Magnetic eld: 400 A/m or less (50/60 Hz)
Position: Upright and horizontal to ground
Structure: Not explosion-proof
Usage location: Indoors
Operating altitude: 2000 m or less
Installation category: II (per CSA22.2 No. 61010-1)
Pollution degree: 2 (per IEC
61010-1 CSA22.2 No. 61010-1)
Transport and Storage Conditions
Environmental conditions during transport and storage
Storage ambient temperature: -25 to 60°C
Storage ambient humidity: 5% to 95% RH (no condensation)
Vibration: 10 to 60 Hz 4.9m/s
2
or less
Shock: 392 m/s
2
or less (packaged condition)
Mechanical Specifications
External dimensions: Approx. 442 × 131 × 159 mm (when six slots are attached)
Weight: Approx. 4.1 kg (maximum total weight)
Installation method: Desktop or oor, panel mount, or attached to a DIN rail
Material: Steel plate, aluminum die-cast, molded plastic resin
Supported Standards
CSA: Conforms to CSA22.2 No. 61010-1, overvoltage category II,
pollution degree 2
UL: Conforms to UL61010B-1 (CSA NRTL/C)
CE: EMC directives; EN61326 Class
A, EN61000-3-2,
EN61000-3-3, EN55011 Class A, Group 1
Low voltage directive: EN61010-1; measurement category II,
pollution degree 2
C-Tick: Conforms to AS/NZS CISPR11 Class A
Group1
Installation category II: An index for dening transient overvoltages (includes the rated
impulse withstand voltage, and applies to electrical equipment
supplied with power from xed installations such as distribution
boards).
Pollution degree 2:
Indicates the degree of adherence by a solid, liquid, or vapor that
reduces the withstand voltage or surface resistance (applies only
to normal indoor atmospheres (with non-conductive pollution).
Measurement category II:
Applies to electrical instruments that are powered by wall outlets
and other xed installations that are wired to distribution boards,
or measurements of such wires.