User Manual

3-9
IM MX100-01E
Troubleshooting and Maintenance
3
2
1
4
5
Index
When calibrating the RTD or resistance range of the 6-CH, Medium-Speed Four-Wire
RTD Resistance Input module
Short the input terminal
of CH3 (connect 0 )
When calibrating the RTD (1 mA)
20 mV range, connect 20 to the
input terminal of CH4.*
* When calibrating the RTD (1 mA) 60 mV range, connect 60 to the input terminal of CH5.
When calibrating the RTD (1 mA) 200 mV range, connect 200
to the input terminal of CH6.
When calibrating the RTD (1 mA) 600 mV range, connect 300
to the input terminal of CH4.
When calibrating the RTD (0.25 mA) 600 mV range, connect 2400
to the input terminal of CH5.
When calibrating the RTD (0.25 mA) 1 V range, connect 3000
to the input terminal of CH6.
Resistance standard
C
A
I
C
A
B
I
B
When calibrating the range of the 4-CH, Medium-Speed Strain Module (-B12, -B35,
and -NDI)
The clamp terminal (-B12, -B35) and NDIS terminal (-NDI) are connected with the
4-gauge method (see below and next page). Resistors R1-R3 use 120 Ω, and a
resistance corresponding to the Zero or Full value is connected to R4.
To correctly calibrate the range, do so in the order Zero, then F
ull.
Calibration Range
Res. R1, R2, and R3 Res. R4 Resistance value
accuracy
Zero Full*
2000 µstrain 120.000 Ω 120.000 Ω 117.154 Ω ±0.005%, ±0.3ppm/°C
20000 µstrain 120.000 Ω 120.000 Ω 113.010 Ω ±0.005%, ±0.3ppm/°C
200000 µstrain 120.000 Ω 120.000 Ω 80.000 Ω ±0.005%, ±0.3ppm/°C
* The Full calibration value of 2000 µstrain range is approximately to 12000 µstrain
The Full calibration value of 20000 µstrain range is approximately to 30000 µstrain
The Full calibration value of 200000 µstrain range is approximately 200000 µstrain
Setting switch (CH1 to CH4)
ON
OFF
No.1
No.2
No.3
No.4
No.5
For -B12, -B35
No.1
OFF
No.2
OFF
No.3
OFF
No.4
OFF
No.5
ON
Connect CH2
only
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
A(+)
B(L)
C(-V)
D(H)
R2
R4
R3
R1
3.3 Calibration