On the MX100, the integral time is determined by the measurement interval setting as
shown in the table below.
4-CH, High-Speed Universal Input Module
Measurement
Interval
Integration
Time
Filter Rejected Noise and Notes
10 ms 1.67 ms
Rectangular
600 Hz and its integer multiples*
50 ms
16.67 ms 60 Hz and its integer multiples
20 ms 50 Hz and its integer multiples
Auto
Automatically detects the power supply frequency
and set 16.67 or 20 ms
100, 200 ms 36.67 ms Trapezoidal 50 Hz or 60 Hz and their integer multiples
500 ms 100 ms Rectangular 10 Hz and its integer multiples
1, 2, 5,10, 20,
30, 60 s
200 ms Cos Fc = 5-Hz low-pass filter
* Since the power supply frequency noise is not rejected, the measured vales may fluctuate
especially with temperature measurement. In such cases, set the measurement interval to 50
ms or higher.
10-CH, Middle-Speed Universal Input Module/6-CH, Medium-Speed Four-Wire RTD
Resistance Input Module
Measurement
Interval
Integration
Time
Filter Rejected Noise and Notes
100, 200 ms 1.67 ms
Rectangular
600 Hz and its integer multiples*
500 ms
16.67 ms 60 Hz and its integer multiples
20 ms 50 Hz and its integer multiples
Auto
Automatically detects the power supply frequency
and set 16.67 or 20 ms
1 s 36.67 ms Trapezoidal 50 Hz or 60 Hz and their integer multiples
2 s 100 ms Rectangular 10 Hz and its integer multiples
5, 10, 20, 30, 60 s
200 ms Cos Fc = 5-Hz low-pass filter
* Since the power supply frequency noise is not rejected, the measured vales may fluctuate
especially with temperature measurement. In such cases, set the measurement interval to 500
ms or higher, or use the 4-CH, High-Speed Universal Input Module
30-CH, Middle-Speed DCV/TC/DI Input Module
Measurement
Interval
Integration
Time
Filter Rejected Noise and Notes
500 ms 1.67 ms
Rectangular
600 Hz and its integer multiples*
1 s
16.67 ms 60 Hz and its integer multiples
20 ms 50 Hz and its integer multiples
Auto
Automatically detects the power supply frequency
and set 16.67 or 20 ms
2 s 36.67 ms Trapezoidal 50 Hz or 60 Hz and their integer multiples
5, 10, 20, 30, 60 s
100 ms Rectangular 10 Hz and its integer multiples
* Since the power supply frequency noise is not rejected, the measured vales may fluctuate
especially with temperature measurement. In such cases, set the measurement interval to 1 s
or higher, or use the 4-CH, High-Speed Universal Input Module or the 10-CH, Medium-Speed
Universal Input Module.
4-CH, Medium-Speed Strain Input Module
Measurement
Interval
Integration
Time
Filter Rejected Noise and Notes
100 ms 1.67 ms
Rectangular
600 Hz and its integer multiples*
200 ms
16.67 ms 60 Hz and its integer multiples
20 ms 50 Hz and its integer multiples
Auto
Automatically detects the power supply frequency
and set 16.67 or 20 ms
500 ms 36.67 ms Trapezoidal 50 Hz or 60 Hz and their integer multiples
1 s 100 ms Rectangular 10 Hz and its integer multiples
2, 5,10, 20,
30, 60 s
200 ms Cos Fc = 5-Hz low-pass filter
* Since the power supply frequency noise is not rejected, the measured vales may fluctuate. In
such cases, set the measurement interval to 200 ms or higher.